Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-168
2450-901-01 Rev. B/September 2013
:TRIGger:TIMer<n>:DELay
These commands set and read the timer delay.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
10e-6 (10
µ
s)
Usage
TRIGger:TIMer<n>:DELay <interval>
TRIGger:TIMer<n>:DELay?
<n>
Trigger timer number (1 to 4)
<interval>
Delay interval in seconds
Details
Each time the timer is triggered, it uses this delay period.
Reading this command returns the delay interval that will be used the next time the timer is triggered.
Example
TRIG:TIM2:DEL 50E-6
Set trigger timer 2 to delay for 50 µs.
Also see