Section 8: TSP command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
8-86
2450-901-01 Rev. B/September 2013
smu.interlock.tripped
This attribute indicates that the interlock has been tripped.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
interlockStatus
= smu.interlock.tripped
interlockStatus
The status of the interlock:
•
smu.OFF
: The interlock is not asserted and the 200 V range is disabled;
lower voltage ranges are available
•
smu.ON
: The interlock signal is asserted and all voltage ranges are available
Details
This command gives you the status of the interlock. When the safety interlock signal is asserted, all voltage
ranges of the instrument are available. However, when the safety interlock signal is not asserted, the 200 V
range is disabled, limiting the nominal output to ±37 V.
When the interlock is not asserted:
•
The front-panel INTERLOCK indicator is on.
•
High voltage ranges are disabled.
•
An event message is generated when you attempt to turn on the source with a voltage higher
than ±20 V.
Example
print(smu.interlock.tripped)
If the interlock is not asserted, returns
smu.OFF
.
If the interlock is asserted, returns
smu.ON
.
Also see
None
smu.measure.autorange
This attribute determines if the measurement range is set manually or automatically for the selected measurement
function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.ON
Usage
autoRange
= smu.measure.autorange
smu.measure.autorange =
autoRange
autoRange
Set the measurement range manually:
smu.OFF
Set the measurement range automatically:
smu.ON