:TRACe:LOG:STATe
This command indicates whether the reading buffer should log informational events.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
defbuffer1: 1 (ON)
defbuffer2: 1 (ON)
User-created buffer: 0 (OFF)
Usage
TRACe:LOG:STATe <logState>, <bufferName>
TRACe:LOG:STATe? <bufferName>
<
logState
>
Disable logging:
OFF
or
0
Enable logging:
ON
or
1
<
bufferNam
e>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified, this
parameter defaults to
defbuffer1
Details
This command indicates whether the reading buffer should log informational events.
Example
TRACe:LOG:STATe?
Query the log state of
defbuffer1
.
Output:
1
Indicates that the log state is on.
TRACe:MAKE "testData", 100
Create a user-defined buffer named
testData
TRACe:LOG:STATe? "testData"
Query the log state of
testData
.
Output:
0
The output indicates that the log state is off.
TRACe:LOG:STATe 1, "testData"
Change the
testData
log state to on.
TRACe:LOG:STATe? "testData"
Query the log state of
testData
.
Output:
1