Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-17
:CALCulate2:<function>:LIMit<Y>:STATe
This command enables or disables a limit test.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
:CALCulate2:<function>:LIMit<Y>:STATe <b>
:CALCulate2:<function>:LIMit<Y>:STATe?
<function>
The function to which this setting applies:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<Y>
The limit that the setting applies to, 1 or 2
<b>
Enable the limit test:
1
or
ON
Disable the limit test:
0
or
OFF
Details
This command enables or disables a limit test for the selected measurement function.
If you send this command without the
<function>
parameter, it will set the state of math operations
for all functions.
Example
CALC2:VOLT:LIM1:LOW 0.25
CALC2:VOLT:LIM1:UPP 2.5
CALC2:VOLT:LIMIT1:STAT ON
CALC2:VOLT:LIMIT1:FAIL?
Set lower limit 1 for voltage to 0.25 volts.
Set upper limit 1 for voltage to 2.5 volts.
Enable limit 1 testing for voltage.
Check the test results.
Also see
None