Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-114
2450-901-01 Rev. B/September 2013
:TRACe:MAKE
This command creates a user-defined reading buffer.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Saved setup
Not applicable
Usage
:TRACe:MAKE <bufferName>, <bufferSize>
<bufferName>
A user-supplied string that indicates the name of a user-defined buffer
<bufferSize>
A number that indicates the maximum number of readings that can be stored in
<bufferName>
; minimum is 10
Details
This command creates a user-defined reading buffer.
You cannot assign user-defined reading buffers the same name as an existing buffer, including
defbuffer1
or
defbuffer2
.
Example
TRACe:MAKE "capTrace", 200
This example creates a 200-element reading buffer
named
capTrace
.
Also see
(on page 3-10)
(on page 3-27)