Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-225
Details
The trigger event detector enters the detected state when an event is detected. When this command is sent, the
instrument does the following actions:
•
Clears the trigger event detector
•
Discards the history of the trigger line
•
Clears the
trigger.tsplinkin[
N
].overrun
attribute
Example
tsplink.line[2].mode = tsplink.MODE_TRIGGER_OPEN_DRAIN
trigger.tsplinkin[2].clear()
Clears the trigger event
on TSP-Link line 2.
Also see
(on page 8-226)
(on page 8-232)
trigger.tsplinkin[N].edge
This attribute indicates which trigger edge controls the trigger event detector for a trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
TSP-Link line
N
reset
Configuration script
trigger.EDGE_FALLING
Usage
detectedEdge
= trigger.tsplinkin[
N
].edge
trigger.tsplinkin[
N
].edge =
detectedEdge
detectedEdge
The trigger mode:
•
Detect falling-edge triggers as inputs:
trigger.EDGE_FALLING
•
Detect rising-edge triggers as inputs:
trigger.EDGE_RISING
•
Detect either falling or rising-edge triggers as inputs:
trigger.EDGE_EITHER
N
The trigger line (1 to 3)
Details
When the edge is detected, the instrument asserts a TTL-low pulse for the output.
The output state of the I/O line is controlled by the trigger logic, and the user-specified output state of the line is
ignored.
Example
tsplink.line[3].mode = tsplink.MODE_TRIGGER_OPEN_DRAIN
trigger.tsplinkin[3].edge = trigger.EDGE_RISING
Sets synchronization line 3
to detect rising edge
triggers as input.
Also see
(on page 8-46)
(on page 8-232)