Section 8: TSP command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
8-46
2450-901-01 Rev. B/September 2013
Example
data = digio.readport()
print(data)
Assume lines 2, 4, and 6 are set high when the
I/O port is read.
Output:
42
This is binary 10101010
Also see
(on page 8-46)
(on page 3-62)
digio.writeport()
This function writes to all digital I/O lines.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.writeport(
data
)
data
The value to write to the port (0 to 63)
Details
The binary representation of the value indicates the output pattern to be written to the I/O port. For example, a
value of 63 has a binary equivalent of 111111 (all lines are set high); a
data
value of 42 has a binary equivalent
of 101010 (lines 2, 4, and 6 are set high, and the other 3 lines are set low).
An instrument reset does not affect the present states of the digital I/O lines.
Example
digio.writeport(63)
Sets digital I/O lines 1 through 6 high (binary
111111).
Also see
(on page 8-45)
(on page 3-62)