Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-215
trigger.clear()
This function clears the trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.clear()
Details
The trigger event detector indicates if an event has been detected since the last
trigger.wait()
command
was sent.
trigger.clear()
clears the trigger event detector and discards the history of trigger events.
Also see
(on page 8-215)
trigger.wait()
This function waits for a trigger event.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
triggered
= trigger.wait(
timeout
)
triggered
true
: A trigger was detected during the timeout period
false
: No triggers were detected during the timeout period
timeout
Maximum amount of time in seconds to wait for the trigger
Details
This function waits up to
timeout
seconds for a trigger on the active command interface. A command interface
trigger occurs when:
•
A GPIB GET command is detected (GPIB only)
•
A VXI-11 device_trigger method is invoked (VXI-11 only)
•
A *TRG message is received
If one or more of these trigger events were previously detected, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Example
triggered = trigger.wait(10)
print(triggered)
Waits up to 10 seconds for a trigger.
If
false
is returned, no trigger was detected
during the 10-second timeout.
If
true
is returned, a trigger was detected.
Also see
(on page 8-215)