Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-217
trigger.timer[N].delay
This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger timer
N
reset
Configuration script
10e-6 (10
µ
s)
Usage
interval
= trigger.timer[
N
].delay
trigger.timer[
N
].delay =
interval
interval
Delay interval in seconds (5.00e-07 to 100,000)
N
Trigger timer number (1 to 4)
Details
Once the timer is enabled, each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[
N
].delaylist = {
interval
}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
This command should not be used with the trigger model.
Example
trigger.timer[1].delay = 50e-6
Set the trigger timer 1 to delay for
50 µs.
Also see
(on page 8-220)
trigger.timer[N].delaylist
This attribute sets an array of timer intervals.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger timer
N
reset
Configuration script
{10e-6}
Usage
intervals
= trigger.timer[
N
].delaylist
trigger.timer[
N
].delaylist =
intervals
intervals
Table of delay intervals in seconds
N
Trigger timer number (1 to 4)