• General parameters: the general parameters include HR, SpO
2
, SpO
2
_PR, SYS, DIA, MAP, NIBP_PR, RR,
S-T, Temp1, Temp2 and TD.
- HR: turn on or off over-limits alarms of HR, set the high and low alarm limits, and set the alarm level. Set-
ting range See
below
Section. 3 alarm level for options: high, medium and low.
The operation for other parameters is similar to that of HR, and we will not cover it again.
- Previous page: click it to turn to previous page.
- Next page: click it to turn to next page.
- All on: select it to enable alarm function for all parameters. Password is required for this operation.
- All off: select it to disable alarm function for all parameters. Password is required for this operation, and
it’s not recommended to disable the alarm function completely.
Notes: 1.
If the parameter alarm switch is set to “OFF”, then icon “
” will be displayed on the correspond-
ing parameter panel. If the parameter panel has two or more related parameters, only all alarm switch for
these related parameters set as “OFF”, icon “
” will be displayed on its parameter panel. For example,
there are SYS, DIA and MAP for NIBP parameter panel, only the alarm switch for SYS, DIA and MAP are set
as “OFF”, then icon “
” will be displayed on the NIBP panel.
2.
The high alarm limit for SpO
2
is fixed to be “100” and is nonadjustable.
• For the monitor configured with IBP, CO
2
, CSM, AG monitoring, then the alarm switch, high/low alarm limits,
and alarm level for these parameters can be set here.
•
ARR:
to set the alarm priority level for 5 arrhythmia (ARR) events, which include ECG Brady, ECG Tachy,
ECG Arrest, VE Run and SVE Run. 3 priority alarm level for option: high, medium and low. The alarm switch
for these 5 ARR events are fixed to be “On” and are nonadjustable. Buttons “All On” and “All Off” are non-
adjustable for the current ARR alarm setting.
•
Others:
to set the alarm volume and timeout setting for apnea alarm.
-
Alarm volume:
to set the alarm sound volume for the monitor. The setting range is “1~10” level, and the
default is 5. Note: for the monitor optioned with “Mute for Spot Check” function, the setting option is 1~10
and “Spot Check”.
-
Apnea:
to set the timeout setting for apnea alarm (in seconds). The options are: off, 10, 15, 20, 25, 30, 35
and 40. Details see Section “
Changing RESP Settings
”.
6.4.1 High and Low Alarm Setting Range
Select “Alarm”
“General Parameter”.
Parameter
Setting range
High limit
Low limit
HR (bpm)
(Low limit+1)~350
0~(High limit-1)
S-T (mV)
(Low limit+0.01)~2.50
-2.5~(High limit-0.01)
SpO
2
(%)
(Low limit+1)~100
0~(High limit-1)
SpO
2
_PR (bpm)
(Low limit+1)~300
0~(High limit-1)
RR (rpm)
(Low limit+1)~150
0~(High limit-1)
TEMP1 (°C)
(Low limit+0.1)~60.0
0~(High limit-0.1)
TEMP2 (°C)
(Low limit+0.1)~60.0
0~(High limit-0.1)
TD (°C)
0.0~5.0
NIBP (Unit) mmHg
Adult
Pediatric
Neonate
SYS
High limit
(Low limit+1) ~280
(Low limit+1) ~200
(Low limit+1) ~135
Low limit
29~ (High limit-1)
29~ (High limit-1)
29~ (High limit-1)
MAP
High limit
(Low limit+1) ~242
(Low limit+1) ~165
(Low limit+1) ~110
Low limit
20~ (High limit-1)
20~ (High limit-1)
20~ (High limit-1)
DIA
High limit
(Low limit+1) ~232
(Low limit+1) ~150
(Low limit+1) ~100
Low limit
10~ (High limit-1)
10~ (High limit-1)
10~ (High limit-1)
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......