Default Alarming Values and Setting Range
Electromagnetic Immunity
The IRMA probe is intended for use in the electromagnetic environment specified below. The customer or the
user of the IRMA probe should assure that it is used in such an environment.
Mode
Parameter
Default value
(for all patient type )
Setting range
(for all patient type )
EtCO
2
High limit
6.6%
(Low limit +0.1)~15%
Low limit
3.3%
0~(High limit - 0.1)%
FiCO2
High limit
0.5%
(Low limit +0.1)~15%
Low limit
0%
0~(High limit - 0.1)%
EtN
2
O
High limit
55
(Low limit +1)~100
Low limit
0
0~(High limit - 1)
FiN
2
O
High limit
53
(Low limit +1)~100
Low limit
0
0~(High limit - 1)
EtAA1
EtHal/Enf/Iso
High limit
3.0%
High limit setting range:
(Low limit +0.1)~15%
Low limit setting range:
0~(High limit - 0.1)%
Low limit
0%
EtSev
High limit
6.0%
EtAA2
Low limit
0%
EtDes
High limit
8.0%
Low limit
0%
FiAA1
EtHal/Enf/Iso
High limit
2.0%
High limit setting range:
(Low limit +0.1)~15%
Low limit setting range:
0~(High limit - 0.1)%
Low limit
0%
EtSev
High limit
5.0%
FiAA2
Low limit
0%
EtDes
High limit
6.0%
Low limit
0%
Immunity test
IEC 61001 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±6 kV air
±6 kV contact
±6 kV air
Floor should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
3A/m
3A/m
Mains power quality should be that of a typical
commercial or hospital environment.
RF emissions
GISPR 11
N/A
Group 1
Mains power quality should be that of a typical
commercial or hospital environment. If the use
of the IRMA probe requires continued operation
during power mains interruptions, it is recom-
mended that the IRMA probe is powered from
an uninterruptible power supply or a battery.
RF emissions
GISPR 11
N/A
Group B
The IRMA probe uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......