61
Improper use of a defibrillator may cause injury to the patient. The user should determine whether to per-
form defibrillation or not according to the patient’s condition.
Before defibrillation, the user must ensure both defibrillator and monitor have passed the system test and
can be safely used jointly.
Do Not use the damaged pressure tube and transducer.
If air bubble appears in pressure tube, please fill the tube with saline solution again. Air bubble may cause
inaccurate reading.
When doing ICP measurement to the patient who is sitting, please keep the pressure transducer and the
top of patient’s ear at the same level, or it may lead to error reading.
When unplugging the cable from the monitor, be sure to hold the head of the connector and pull it out.
Each time when connecting transducer kit or using a new transducer kit, zero calibration to the IBP trans-
ducer must be carried out.
Before using the tube, cable and/or transducer, make sure that all accessories meet the performance re-
quirements which is not changed by aging or environmental conditions.
12.3 Setting Up the IBP Measurement
If your monitor is configured with Plugin IBP Module, then you need to make sure that the IBP Plug-and-Play
box is securely inserted. If the box is inserted successfully, then the IBP icon appears on the upper right side
of the screen. Refer to the following description.
If your monitor is configured with internal IBP Module, then the IBP icon appears on the upper right side of the
screen.
12.3.1 Understanding IBP Icon and IBP Plugin Box (Optional)
The Plugin box (optional) can be configured with 2 IBP module and one C.O. module, or be configured with 2
IBP modules, as shown in below figure, or be configured with C.O. module, refer to the monitor in your hand.
For some models of the device, the icon for the IBP plugin box will be displayed on the upper right corner of
the monitor, for example, “
” for IBP and C.O. Plugin box, “
” for IBP Plugin box, “
” for C.O. Plugin box.
Appearance for IBP and C.O. Plugin box
The icons “IBP1” and “IBP2” are the connectors to the IBP1 and IBP2. If the Plugin box is inserted into the left
slot, then “IBPL1” and “IBPL2” will be displayed on the screen of the monitor. The connector will be different
because of the different configuration, please refer to the monitor in your hand.
“IBP1” and “IBP2” marked on the device is the icons for IBP cuff connector of the internal IBP module. The
internal IBP module will be shown as “IBP1” and “IBP2” in software and operation.
12.3.2 IBP Transducer Kit Connection
1. On the signal input panel of the patient monitor, there are 2 (or 4) connectors with label “IBP1” and “IBP2”
(or label “IBP1”, “IBP2” , “IBP3”, and “IBP4”) depending on your monitor configuration, which are used to
connect the IBP transducer kit. If you want to use IBP monitoring function, please connect the IBP trans-
ducer kits into the IBP connectors on the panel of the patient monitor. Please make sure the cables of the
transducer kits are not folded or twisted.
2. Prepare the pressure tube and transducer by filling the pressure tube with saline solution, and make sure
that the pressure tube is free from any air bubble.
3. Connect the pressure catheter coming from the patient to the pressure tube of the transducer kit, and en-
sure there is no air present in the pressure tube, pressure transducer or pressure catheter.
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......