The device is not intended for use in an oxygen rich environment.
The surface of the device can be cleaned and disinfected with 75% Ethanol, no need to sterilize.
26.16 Operating Environment
1. Ambient temperature range: 5°C~40°C
Relative humidity: 15%~85%, non-condensing
Atmospheric pressure: 70kPa ~106.0kPa
Power Voltage: (100-240)VAC
Power frequency: 50Hz/60Hz
2. This equipment should be situated in a place protected against direct sunlight, so as to prevent overheating
inside the equipment.
3. The device should be stored and used within specified temperature, humidity and atmospheric pressure
range, or it may cause damage to the device or inaccurate measurement result.
4. If the device gets wet by accident, the operator should NOT power it on directly until it has been air-dry
enough to avoid any damage to it.
5. Do not use this equipment in an environment with toxic or inflammable gas.
6. This equipment should be placed on a stand or flat platforms, so as to prevent possible shock.
7. Do not use this equipment in combination with any equipment other than those expressly permitted in the
manual.
8. The monitor is defibrillator discharge proof and can be used with electrosurgical unit. But when the device
is used together with defibrillator or electrosurgical equipment, the user (doctor or nurse) should keep the
patient under close surveillance for his/her safety. Refer to the following function description for specific
protective measures or notes.
9. Make sure that the equipotential grounding terminal is grounded correctly.
10. Do not use mobile phone nearby, so as to avoid strong radiant field interference.
26.17 Storage
If the equipment will not be used for long period of time, wipe it clean and keep it in the packaging, which shall
be kept in a dry and good ventilation place free from dust and corrosive gases.
Storage environment: ambient temperature: -20~60°C
relative humidity:
10%~95%
atmospheric pressure: 53kPa~106kPa
26.18 Transportation
This monitor should be transported by land (vehicle or railway) or air in accordance with the contractual terms.
Do not hit or drop it with force.
Transportation environment: ambient temperature: -20~60°C
relative humidity:
10%~95%
atmospheric pressure: 53kPa~106kPa
26.19 Packaging
The product is packed in high quality corrugated cartons with foam inside to protect the apparatus against
damage in the handling process.
Gross weight: Details see the indication on the outer package.
Dimension: Details see the indication on the outer package.
Chapter 27
Troubleshooting
Note: In case of trouble of this machine in service, follow the instructions below to eliminate the problem first.
If the attempt fails, contact the dealer in your local area or the manufacturer.
Do NOT open the monitor cabinet without permission
During operation, if the system error and the prompt pops up on the screen, then the system will shutdown,
and it restarts within 5 seconds and resume to factory default settings.
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......