1. The surgeon places the catheter into the patient artery. Based on the monitoring of IBP or X-ray scanning,
the surgeon can determine the target position of the catheter.
2. On “Menu” screen, rotate the Navigator Knob and move the focus on the icon “Cardiac Output”, press the
knob to enter C.O. measuring screen.
3. Set the parameters according to the type of catheter and the injecting condition.
4. Press “Start” button, and drop one injectate, then the measured result will be displayed. Take C.O. meas-
urement in multiple times, and get the average result.
5. Obtain other hemodynamic calculations.
14.4 Changing C.O. Settings
Preparing:
Before make measurement, please setup the corresponding parameters as following.
Select “Menu”
“C.O.”
“Parameter” to enter into C.O. parameter related settings.
Floating Catheter:
•
Brand
: brand of the catheter.
•
Model
: model of the catheter.
•
Computing constant
: setting the correction constant according to the catheter’s brand and model.
Injectate:
•
Temp. Examin. Mode
: temperature measuring method for the injectate. 2 options: Manual and Auto.
•
Temperature
: the temperature value can be input when the Temp. Examin. Mode is “Manual”.
•
Volume
: injectate volume.
14.5 Understanding
the C.O. Display
Select “Menu”
“C.O.” to enter into C.O.
display.
Display screen for C.O. Measurement
C.O. record list area
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......