5. With the oscillometric method, when blood pressure is measured, the inflation pressure of the cuff will be
automatically adjusted according to the previous measurement. Generally, the initial inflation pressure is
150mmHg (for the adult mode) or 120mmHg (for pediatric) or 70 mmHg (for neonate) when it is powered on.
Following that, 28mmHg (for adult) or 25mmHg (for pediatric) or 25mmHg (for the neonate) will be added
on the basis of the last measurement of systolic pressure. In this way, when the blood pressure rises or the
patient is changed, the device may fail in giving the result after the first-time inflation. This device will auto-
matically adjust the inflation pressure until the measurement is taken, after that, up to four times retry will be
allowed.
6. When an adult patient is monitored, the device may fail in giving the blood pressure measurement if the
Pediatric or neonate patient type is selected.
7. When taking NIBP measurement on pediatric or neonatal patients, the operator must select correct patient
type depending on different patients (refer to NIBP menu setup) and do NOT operate with the adult patient
type setting. The high inflation pressure for adult is not suitable for pediatric patients.
9.6 Understanding the NIBP Numerics
NIBP Panel:
• “
NIBP
”: the label of blood pressure. “123” is the systolic pressure value, “80” is the diastolic pressure value,
and “95” is the mean arterial pressure.
• “
mmHg
”: the unit of blood pressure value, 1kPa = 7.5mmHg.
• “
PR 62
”: pulse rate value when taking blood pressure measurement.
• “
Manu
”: the icon of NIBP measurement mode. There are 3 modes: “Manual”, “Auto” and “STAT”. When in
“AUTO” mode, a count-down timer is displayed as well.
9.7 Changing NIBP Settings
Select “Menu”
“NIBP” to enter into NIBP related setting.
•
Mode
: “MANU”, “AUTO”, “STAT” and “Customized Multi-cycle” can be selected. The default is “MANU”.
- In “MANU” mode, press the NIBP measure key “
” manually can start or stop NIBP measure-
ment.
- In “AUTO” mode, the device repeats NIBP measurement automatically with the set time interval. In this
mode, the manual interference still works.
- In “STAT” mode (only used for adult), press the NIBP measure key “
”, the device will do NIBP
measurement again and again. The device will not stop making measurement until the measuring time is
over 5 minutes or the operator stops it manually.
•
Customized Multi-cycle
: move the cursor on “Customized Multi-cycle” and click “OK” to enable this func-
tion. The user can customize the related parameters: Phase, time cycle (the time interval between two
measurements) and repeats. There are 5 phases: A, B, C, D and E. The user can set the time cycle and
repeats for Phase A to Phase E.
•
Time cycle
: 1min, 2min, 3min, 4min, 5min, 10min, 15min, 20min, 25min, 30min, 35min, 40min, 45min,
50min, 55min, 1h, 1.5h, 2h, 2.5h, 3h, 3.5h and 4h for optional.
•
Repeats
: OFF, 1, 2, ...9 and 10 for optional.
For example, firstly, the monitor enter into Phase A (making NIBP measurement once every 5 minutes, and
repeat once only); Secondly, entering into Phase B (making NIBP measurement once every 10 minutes,
and repeat once only); Thirdly, entering into Phase C (making NIBP measurement once every 20 minutes,
and repeat 2 times); Fourthly, entering into Phase D (making NIBP measurement once every 30 minutes,
and repeat 5 times); Lastly, entering into Phase E (making NIBP measurement once every 60 minutes, and
repeat once only). In the period of 30 minutes, if the NIBP measurement is less than 6 times and NIBP meas-
urement mode is not changed, then the monitor will start to make NIBP measurement from Phase A to E
automatically.
NIBP label and unit
Systolic pressure value
NIBP measurement mode
Diastolic pressure value
Mean arterial pressure
Pulse rate value
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......