109
Mode 1: Automatic inflation for the pressure accuracy verification
At this mode, the monitor can activate the inflation, so the pressure will increase automatically until it exceeds
the limit value specified in table A. This pressure limit value depends on the patient type selection as shown
in table A:
Table A
During the inflation, the Monitor will close the deflating valve, and the pressure value will be shown during the
process. If there is no manual deflation operation, the pressure will persist until deflation by manual operation,
so it is necessary to use a manual valve for doing adequate deflation in several steps to verify the pressure
accuracy in the full scale of measurement range.
Mode 2: Manual inflation for the pressure accuracy verification
At this mode, the pressure should be increased manually by a pumping balloon, and the verification can be
done by applying different pressure value manually. If the increased pressure exceeds the given limit as shown
in table B, the Monitor will deflate automatically because of over-pressure protection.
Table B
Module
testing
software
Air tube
NIBP cuff with dual air tube
This belongs to the monitor
Manual valve
Air vent
Air tube
Air tube
Inflation
balloon
Increase the pressure
manually through the
inflatable balloon
NIBP
module
Mercury blood
pressure meter
Air tube
Adult
240mmHg
Pediatric
200mmHg
Neonate
120mmHg
Adult
300mmHg
Pediatric
240mmHg
Neonate
140mmHg
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......