6.2 Alarm Levels
By severity, the monitor’s alarm can be classified into three categories: high level, medium level and low level.
In addition, the monitor has preset alarm levels about physiological alarm and technical alarm.
Related to 3 levels alarm sources, medical and nursing staff should have different response to deal with poten-
tial dangers, the detailed demands are as follows:
1. High priority alarm: medical and nursing staff should response immediately.
2. Medium priority alarm: medical and nursing staff should response quickly.
3. Low priority alarm: medical and nursing staff should response as soon as possible.
NOTE:
Some models may have only Medium and Low alarm priority because of the different configuration.
Physiological alarm
Alarm priority level Alarm Source
Alarm Event
High
ECG
Unable to detect HR, ECG Arrest, ECG Brady, ECG Tachy, VE
RUN, SVE RUN, HR is too high, HR is too low, S-T is too high,
S-T is too low, VE Run, ECG VPCEST
SpO
2
Unable to detect SpO
2
, SpO
2
too high, SpO
2
too low, PR too
high, PR too low
Respiration
Apnea, RR too high, RR too low
Temperature
Temp1 too high, Temp1 too low, Temp2 too high, Temp2 too low,
TD too high
NIBP
NIBP SYS too high, NIBP SYS too low, NIBP DIA too high, NIBP
DIA too low, NIBP MAP too high, NIBP MAP too low, PR too
high, PR too low
CO
2
RR too high, RR too low, EtCO
2
too high, EtCO
2
too low, InsCO
2
too high, InsCO
2
too low
CSM
CSI is too high, CSI is too low
Medium
ECG
VE RonT, SVE RonT, S-T1 too high, S-T1 too low, S-T2 too high,
S-T2 too low, S-T3 too high, S-T3 too low, S-T4 too high, S-T4
too low, S-T5 too high, S-T5 too low, S-T6 too high, S-T6 too
low, S-T7 too high, S-T7 too low
Low
ECG
Miss Beat, VE Early, SVE Early, VE Couplet, SVE Couplet, VE
Short Run, SVE Short Run, SVE Run, VE Insert, SVE Insert, VE
Bigeminy, SVE Bigeminy, VE Trigeminy, SVE Trigeminy, Multiform
Beat
Technical alarm
Alarm priority level Alarm source
Alarm Events
High
System
Low battery, SpO
2
modular malfunction, Unknown error
CO
2
CO
2
sensor failed, CO
2
sensor temperature is too high
Medium
ECG/SpO
2
Lead(s) off, SpO2 Probe off, SpO
2
value exceeds measuring
range, HR exceeds measuring range
CO
2
CO
2
sensor off, need to perform zero resetting, CO
2
over de-
tect value, check air adapter, check sampling tube, CO
2
zeroing
failed.
Zeroing information (like “Zeroing starting”, “Zeroing success-
ful”) appears on the window
IBP
IBP1 Probe off, IBP2 Probe off, IBP3 Probe off, IBP4 Probe off
Low
NIBP
Self-test failed, System error, Over time, Signal weak, Cuff error,
Air leakage, Pressure error, Out of range, Too much motion, Over
pressure, Signal saturation, Detected air leakage, BP exceeds
measuring range,
TEMP
Temperature self-detect failed, TEMP1/TEMP2 exceed(s) meas-
uring range
CSM
Electrodes off
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......