37
6.3 Alarm Indicators
When an alarm occurs, the monitor will indicate it to the user through visual or audible alarm indications.
6.3.1 Alarm lamp
6.3.2 Alarm message
When an alarm occurs, an alarm message will appear in the technical or physiological alarm area. And the
alarm message is shown in different color to match the alarm level (high priority alarm in red, medium priority
alarm in yellow, and low priority alarm in white).
6.3.3 Highlighting Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will be high-
lighted.
6.3.4 Audible Alarm Tones
The alarm tone is distinct from the heart beat tone, keystroke tone and pulse tone in frequency. The alarm
tones identify the alarm levels as follows:
• High priority alarm: beep+beep+beep+pause+beep+beep+beep
• Medium priority alarm: triple beep
• Low priority alarm: single beep
When multiple alarms of different levels occur simultaneously, the monitor will select the alarm of the highest
level and give visual and audible alarm indications accordingly.
6.3.5 Alarm Status Symbols:
Apart from the aforementioned alarm indicators, the monitor still uses the following symbols telling the alarm
status.
• For the customer purchased the monitor with configuration of “Alarm pause”, then this area shows the alarm
sound status. There are 3 alarm sound status:
1 “
” means the alarm sound is enabled;
2 “
” means the alarm sound is disabled or means the alarm volume is 0 (alarm is mute), at this time the user
should pay more attention to the patient;
3 “
” means the alarm sound is paused. Text display area: when the alarm sound is paused, it dis-
plays counting-down time, otherwise it displays the date. Move the cursor on this text display area can bring
up the edit box of “Alarm volume” setting.
• For the customer purchased the monitor with configuration of “Alarm Reset”, then this area shows the alarm
sound status. There are 2 alarm sound status:
1 “
” means the alarm system keeps on (is on “Alarm activating status”), and when alarm event occurs,
it displays icon “
”;
2 when alarm event occurs, after pressing alarm reset key, then it displays icon “
”. Move the cursor on
this icon can bring up the edit box of “Alarm volume” setting.
Remark:
•
Alarm reset key: pressing this key can perform alarm reset for the current activated alarm event (that’s
to say, the audible alarming will be off, but the visual alarming still keeps effective), then icon “
”
displays on the upper right corner of the screen. The monitor can response to a new alarm event during the
alarm reset status, that’s to say, both visual and audible alarming will be effective when there is a new alarm
Lamp color
Alarm level
Alarm event
Red flashing
High priority alarm
ECG Brady, ECG Tachy, Low battery, Vital sign alarm
Yellow flashing
Medium priority alarm
Lead off, Probe off, Sensor off, VE RonT, SVE RonT
Yellow
Low priority alarm
Other Arrhythmia event
Green
Normal
Summary of Contents for 35145
Page 12: ...12 SERIE III 11 Vista anteriore Vista posteriore Lato sinistro Lato destro...
Page 13: ...13 SERIE IV 11 Vista anteriore Vista posteriore Modulo esteso Lato sinistro Lato destro...
Page 135: ...135...
Page 136: ......
Page 148: ...12 SERIES III 11 Front view Rear view Left view Right view...
Page 149: ...13 SERIES IV 11 Front view Rear view Extended Module Left view Right view...
Page 267: ...131...
Page 268: ......