Operating Instructions

Proline Prowirl F 200

PROFIBUS PA

Vortex flowmeter

BA01690D/06/EN/01.18

71384786
Valid as of version

01.02.zz (Device firmware)

Summary of Contents for Proline Prowirl F 200

Page 1: ...Products Solutions Services Operating Instructions Proline Prowirl F 200 PROFIBUS PA Vortex flowmeter BA01690D 06 EN 01 18 71384786 Valid as of version 01 02 zz Device firmware...

Page 2: ...danger to individuals or the facility read the Basic safety instructions section carefully as well as all other safety instructions in the document that are specific to working procedures The manufact...

Page 3: ...nting position 20 6 1 2 Requirements from environment and process 23 6 1 3 Special mounting instructions 24 6 2 Mounting the measuring device 25 6 2 1 Required tools 25 6 2 2 Preparing the measuring d...

Page 4: ...via write protection switch 119 10 8 Application specific commissioning 120 10 8 1 Steam application 120 10 8 2 Liquid application 121 10 8 3 Gas applications 122 10 8 4 Calculation of the measured v...

Page 5: ...15 1 Device specific accessories 185 15 1 1 For the transmitter 185 15 1 2 For the sensor 186 15 2 Service specific accessories 186 15 3 System components 187 16 Technical data 188 16 1 Application 1...

Page 6: ...njury CAUTION CAUTION This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or medium injury NOTICE NOTE This symbol contains information on procedures an...

Page 7: ...esses or actions that are preferred Forbidden Procedures processes or actions that are forbidden Tip Indicates additional information Reference to documentation A Reference to page Reference to graphi...

Page 8: ...ation Transmitter Brief Operating Instructions Guides you quickly to the 1st measured value Part 2 The Transmitter Brief Operating Instructions are aimed at specialists with responsibility for commiss...

Page 9: ...due to process pressure are labeled accordingly on the nameplate To ensure that the measuring device remains in proper condition for the operation time Keep within the specified pressure and temperatu...

Page 10: ...risk of electric shock gloves must be worn 2 4 Operational safety Risk of injury Operate the device in proper technical condition and fail safe condition only The operator is responsible for interfere...

Page 11: ...access via a password A password can be used to protect against write access to the device parameters This password locks write access to the device parameters via the local display or another operat...

Page 12: ...a mechanical unit Remote version transmitter and sensor are mounted in separate locations 3 1 Product design 1 2 3 4 5 6 7 8 9 E A0020649 1 Important components of a measuring device 1 Electronics co...

Page 13: ...Center Depending on the device version the CD ROM might not be part of the delivery The Technical Documentation is available via the Internet or via the Endress Hauser Operations App see the Product i...

Page 14: ...late Order code Ext ord cd Ser no Date i i Patents 322540 0001 1 2 3 4 5 6 7 8 9 10 11 12 14 15 16 17 13 A0032237 2 Example of a transmitter nameplate 1 Manufacturing location 2 Name of the transmitte...

Page 15: ...13 12 4 5 14 11 A0034423 3 Example of a sensor nameplate 1 Name of the sensor 2 Nominal diameter of sensor 3 Flange nominal diameter nominal pressure 4 Serial number ser no 5 Measuring tube material 6...

Page 16: ...0034161 4 Example of a sensor nameplate 1 Nominal diameter of sensor 2 Flange nominal diameter nominal pressure 3 Measuring tube material 4 Measuring tube material 5 Serial number ser no 6 Maximal per...

Page 17: ...material 15 Seal material 16 Medium temperature range Order code The measuring device is reordered using the order code Extended order code The device type product root and basic specifications manda...

Page 18: ...and dust free place Do not store outdoors Storage temperature 50 to 80 C 58 to 176 F 5 2 Transporting the product Transport the measuring device to the measuring point in the original packaging A00292...

Page 19: ...g a forklift 5 3 Packaging disposal All packaging materials are environmentally friendly and 100 recyclable Measuring device secondary packaging polymer stretch film that conforms to EC Directive 2002...

Page 20: ...zontal orientation transmitter head down A0015590 4 5 D Horizontal orientation transmitter head at side A0015592 4 1 In the case of liquids there should be upward flow in vertical pipes to avoid parti...

Page 21: ...Expansion 7 Control valve 8 Two measuring devices in a row where DN 25 1 directly flange on flange 9 Two measuring devices in a row where DN 40 1 for spacing see graphic If there are several flow dist...

Page 22: ...for steam Example for H2O condensate 80 C p 10 bar abs 965 kg m3 t 240 C 4 39 kg m3 v 2 5 m s v 40 m s p 0 0085 965 2 5 2 51 3 mbar p 0 0085 4 394 39 40 2 59 7 mbar density of the process medium v av...

Page 23: ...o 140 F 1 Sensor Non hazardous area 40 to 85 C 40 to 185 F 1 Ex i Ex nA Ex ec 40 to 85 C 40 to 185 F 1 Ex d 40 to 85 C 40 to 185 F 1 Ex d Ex ia 40 to 85 C 40 to 185 F 1 Local display 40 to 70 C 40 to...

Page 24: ...r Sensor version option CA mass 316L 316L integrated temperature measurement 200 to 400 C 328 to 750 F Order code for Sensor version option CB mass Alloy C22 316L integrated temperature measurement 20...

Page 25: ...uring clamps Allen key3 mm For turning the transmitter housing Open ended wrench8 mm For opening the securing clamps Allen key3 mm For sensor For flanges and other process connections Corresponding mo...

Page 26: ...entries do not point upwards A0029263 6 2 4 Mounting the transmitter of the remote version LCAUTION Ambient temperature too high Danger of electronics overheating and housing deformation Do not exceed...

Page 27: ...To provide easier access to the connection compartment or display module the transmitter housing can be turned max 350 8 mm 8 mm 1 2 3 A0032242 1 Release the fixing screw 2 Turn the housing to the de...

Page 28: ...eassemble the transmitter 6 3 Post installation check Is the device undamaged visual inspection Does the measuring device conform to the measuring point specifications For example Process temperature...

Page 29: ...expected Signal cable Pulse frequency switch output Standard installation cable is sufficient PROFIBUS PA Twisted shielded two wire cable Cable type A is recommended For further information on plannin...

Page 30: ...ble Connecting cable reinforced Cable reinforced 2 2 0 34 mm2 22 AWG PVC cable with common shield 2 pairs pair stranded and additional steel wire braided sheath 1 Flame resistance According to DIN EN...

Page 31: ...ity protection 7 1 5 Pin assignment of device plug 1 2 4 3 Pin Assignment Coding Plug socket 1 PROFIBUS PA A Plug 2 Grounding 3 PROFIBUS PA 4 Not assigned 7 1 6 Shielding and grounding Optimal electro...

Page 32: ...nts connect only one point of the shielding directly to the reference ground 3 In systems without potential equalization the cable shielding of fieldbus systems should be grounded on one side only for...

Page 33: ...tion C Local operation SD02 DC 1 V Order code for Display Operation option E Local operation SD03 with lighting backlighting not used DC 1 V Order code for Display Operation option E Local operation S...

Page 34: ...the transmitter depends on the following order code Electrical connection Option A B C D terminals Option I M device plug Connection via terminals 10 0 4 mm in 20mm 3 mm 1 2 3 4 5 6 A0032239 1 Loosen...

Page 35: ...two terminal holes while simultaneously pulling the cable end out of the terminal 7 2 2 Connecting the remote version LWARNING Risk of damaging the electronic components Connect the sensor and transm...

Page 36: ...M12 device connector is used for connection in the transmitter housing All other approvals Use of connecting cable standard Terminals are always used to connect the connecting cable in the sensor con...

Page 37: ...ide the connecting cable through the cable entry and into the connection housing if using a connecting cable without an M12 device plug use the shorter stripped end of the connecting cable 4 Wire the...

Page 38: ...rtment cover 3 Pull out the display module with a gentle rotational movement To make it easier to access the lock switch attach the display module to the edge of the electronics compartment 8 mm TX 10...

Page 39: ...on board of the wall housing by pressing in the locking clip on the connector Remove the transmitter housing 9 Guide the connecting cable through the cable entry and into the connection housing if usi...

Page 40: ...o 1 7 Nm 13 Reverse the removal procedure to reassemble the transmitter housing 7 2 3 Ensuring potential equalization Requirements Please consider the following to ensure correct measurement Same elec...

Page 41: ...e version are all the device plugs firmly tightened 34 Only for remote version is the sensor connected to the right transmitter Check the serial number on the nameplate of the sensor and transmitter D...

Page 42: ...l F 200 PROFIBUS PA 42 Endress Hauser 8 Operation options 8 1 Overview of operation options 1 2 E ESC 3 A0032227 1 Local operation via display module 2 Computer with operating tool e g FieldCare SIMAT...

Page 43: ...on Diagnostics Access status display Output Operating menu for experts Language Operatation Language Parameter 1 Setup Submenu 1 Submenu n Device tag Advanced setup Enter access code Parameter 1 Param...

Page 44: ...ontains up to 5 currently pending diagnostic messages Event logbook Contains event messages that have occurred Device information Contains information for identifying the device Measured values Contai...

Page 45: ...signals 142 F Failure C Function check S Out of specification M Maintenance required Diagnostic behavior 143 Alarm Warning Locking the device is locked via the hardware Communication communication vi...

Page 46: ...rameter 80 8 3 2 Navigation view In the submenu In the wizard 4 2 1 3 5 Operation 0091 1 Access stat disp Operator Locking status Display A0013993 EN S 4 2 1 5 3 Curr output 1 Assign curr Volume flow...

Page 47: ...gation path in the Operation menu Setup Appears In the menu next to the Setup selection At the left in the navigation path in the Setup menu Diagnostics Appears In the menu next to the Diagnostics sel...

Page 48: ...ls are available in the input mask of the numeric and text editor Numeric editor Symbol Meaning 0 9 Selection of numbers from 0 to 9 Inserts decimal separator at the input position Inserts minus sign...

Page 49: ...ne position to the left Deletes one character immediately to the left of the input position 8 3 4 Operating elements Operating key s Meaning Minus key In a menu submenu Moves the selection bar upwards...

Page 50: ...ou to the next higher level If help text is open closes the help text of the parameter Pressing the key for 2 s returns you to the operational display home position With a Wizard Exits the wizard and...

Page 51: ...Hauser 51 2 Press simultaneously The context menu is closed and the operational display appears Calling up the menu via the context menu 1 Open the context menu 2 Press to navigate to the desired men...

Page 52: ...1 0098 1 0098 1 XXXX XXX X X 10 50 2800 Hz mA 3 4 5 6 7 8 1 2 Display operat Display operat Setup Main menu English Main menu Format display Display Contrast display Display intervall 1 value max Setu...

Page 53: ...erent channel is accessed Enter the direct access code with the corresponding channel number Example Enter 00914 2 Assign process variable parameter For the direct access codes of the individual param...

Page 54: ...ABC 1x 001 FT 10 DEFG HIJK LMNO PQRS TUVW XYZ Aa1 ABC 001 FT 10 DEFG HIJK LMNO PQRS TUVW XYZ Aa1 ABC 001 FT 10 DEFG HIJK ABC A a 1 001 FT 10 DEFG HIJK ABC A a 1 001 FT 10 3456 789 Aa1 012 001 FT 10 34...

Page 55: ...n defined 1 1 The user only has write access after entering the access code Access authorization to parameters Operator user role Access code status Read access Write access After an access code has b...

Page 56: ...ice is restarted To activate the keylock manually 1 The device is in the measured value display Press for at least 2 seconds A context menu appears 2 In the context menu select the Keylock on option T...

Page 57: ...vice Via service interface CDI 1 2 3 E A0034056 1 Service interface CDI Endress Hauser Common Data Interface of the measuring device 2 Commubox FXA291 3 Computer with FieldCare operating tool with COM...

Page 58: ...Operating Instructions BA00027S and BA00059S User interface 6 5 3 2 1 Xxxxxx 7 P P P P 4 8 9 10 11 Xxxxxxx Good Status Device tag Xxxxxxx Device name Mass flow 12 34 kg h Volume flow 12 34 m h Mass f...

Page 59: ...gether with the device type managers DTMs it presents a convenient comprehensive solution For details see Innovation Brochure IN01047S Source for device description files See information 60 8 4 4 SIMA...

Page 60: ...ice descriptions FieldCare www endress com Download Area CD ROM contact Endress Hauser DVD contact Endress Hauser DeviceCare www endress com Download Area CD ROM contact Endress Hauser DVD contact End...

Page 61: ...ow Channel Analog Input 2 mass flow Channel Analog Input 3 corrected volume flow Channel totalizer volume flow The Profile GSD that is to be used is specified in the Ident number selector parameter by...

Page 62: ...he measuring device behaves identically Replacing the measuring devices without changing the GSD file or restarting the controller In the procedure described below the device can be replaced without i...

Page 63: ...ck 1 to 3 To optimize the data throughput rate of the PROFIBUS network it is advisable to only configure modules that are processed in the PROFIBUS master system If this results in gaps between the co...

Page 64: ...sured value floating point number IEEE 754 Status TOTAL module Transmit a totalizer value from the measuring device to the PROFIBUS master Class 1 A selected totalizer value along with the status is c...

Page 65: ...SETTOT and TOTAL functions SETTOT Control the totalizers via the PROFIBUS master TOTAL Transmit totalizer value along with the status to the PROFIBUS master Three Totalizer blocks are available slot...

Page 66: ...Factory setting Value MODETOT meaning Totalizer 1 2 and 3 0 balancing Data structure Output data of SETTOT and MODETOT Byte 1 Byte 2 Control variable 1 SETTOT Control variable 2 MODETOT Input data of...

Page 67: ...asuring device to transmit the state of device functions to the PROFIBUS master Class 1 The DI module cyclically transmits the discrete input value along with the status to the PROFIBUS Master Class 1...

Page 68: ...n block Device function Values control meaning 891 DO 1 Flow override 0 disable device function 1 enable device function 1429 DO 2 Start verification 1 1 Only available with the Heartbeat Verification...

Page 69: ...tartup the local display switches automatically from the startup display to the operational display If nothing appears on the local display or a diagnostic message is displayed refer to the section on...

Page 70: ...tion Setup Diagnostic Operation Setup Main menu English Main menu Medium selection Setup Language A0034189 EN 19 Taking the example of the local display Setup Device tag 71 Medium selection 72 System...

Page 71: ...ag name 1 Tag name Enter the tag name in the FieldCare operating tool 58 Navigation Setup menu Device tag Parameter overview with brief description Parameter Description User entry Factory setting Dev...

Page 72: ...eter Prerequisite Description Selection User entry Factory setting Select medium Select medium type Gas Liquid Steam Steam Select gas type The following conditions are met Order code for Sensor versio...

Page 73: ...integrated temperature measurement In the Select medium parameter the Gas option is selected and in the Select gas type parameter the Natural gas option is selected Select the norm the enthalpy calcu...

Page 74: ...unit 75 Volume unit 75 Mass flow unit 75 Mass unit 75 Corrected volume flow unit 75 Corrected volume unit 75 Pressure unit 75 Temperature unit 75 Energy flow unit 76 Energy unit 76 Calorific value uni...

Page 75: ...volume flow unit Result The selected unit applies for Corrected volume flow parameter 132 Unit choose list Country specific Nm h Sft h Corrected volume unit Select corrected volume unit Unit choose l...

Page 76: ...lorific value unit Result The selected unit applies for Reference gross calorific value Unit choose list Country specific kJ Nm Btu Sft Calorific value unit Mass The following conditions are met Order...

Page 77: ...viscosity unit Select dynamic viscosity unit Result The selected unit applies for Dynamic viscosity parameter gases Dynamic viscosity parameter liquids Unit choose list Pa s Length unit Select length...

Page 78: ...iable Volume flow Mass flow Corrected volume flow Flow velocity Temperature Calculated saturated steam pressure Steam quality Total mass flow Energy flow Heat flow difference Reynolds number Density P...

Page 79: ...es you systematically through all the parameters that can configured for configuring the local display Navigation Setup menu Display Display Format display 80 Value 1 display 80 0 bargraph value 1 80...

Page 80: ...ded Enter 100 value for bar graph display Signed floating point number Depends on country and nominal diameter Value 2 display A local display is provided Select the measured value that is shown on th...

Page 81: ...tion User entry Factory setting Device address Enter device address 0 to 126 126 10 4 7 Configuring the low flow cut off The Low flow cut off wizard systematically guides the user through all the para...

Page 82: ...On value low flow cutoff One of the following options is selected in the Assign process variable parameter 82 Volume flow Corrected volume flow Mass flow Reynolds number Enter on value for low flow c...

Page 83: ...tup Main menu English Main menu Setup Setup Advanced setup Ent access code Advanced setup Def access code Device tag Language Medium selection A0034208 EN The number of submenus can vary depending on...

Page 84: ...alues for the measuring application can be set Navigation Setup menu Advanced setup Medium properties Medium properties Enthalpy type 85 Calorific value type 85 Reference combustion temperature 85 Ref...

Page 85: ...ference combustion temperature parameter is visible Enter reference combustion temperature to calculate the natural gas energy value Dependency The unit is taken from the Temperature unit parameter 20...

Page 86: ...nstant Z for gas under reference conditions 0 1 to 2 1 Linear expansion coefficient The following conditions are met The Liquid option is selected in the Select medium parameter The User specific liqu...

Page 87: ...iscosity Gases The following conditions are met Order code for Sensor version Option Volume or Option Volume high temperature The Gas option or the Steam option is selected in the Select medium parame...

Page 88: ...m properties Gas composition Gas composition Gas type 90 Gas mixture 90 Mol Ar 91 Mol C2H3Cl 91 Mol C2H4 91 Mol C2H6 91 Mol C3H8 92 Mol CH4 92 Mol Cl2 92 Mol CO 92 Mol CO2 93 Mol H2 93 Mol H2O 93 Mol...

Page 89: ...F 200 PROFIBUS PA Commissioning Endress Hauser 89 Mol n C5H12 96 Mol n C6H14 96 Mol n C7H16 96 Mol n C8H18 96 Mol n C9H20 96 Mol Ne 97 Mol NH3 97 Mol O2 97 Mol SO2 97 Mol Xe 97 Mol other gas 98 Relat...

Page 90: ...CO Carbon dioxide CO2 Sulfur dioxide SO2 Hydrogen sulfide H2S Hydrogen chloride HCl Methane CH4 Ethane C2H6 Propane C3H8 Butane C4H10 Ethylene C2H4 Vinyl Chloride C2H3Cl Methane CH4 Gas mixture The f...

Page 91: ...meter the Gas mixture option is selected In the Gas mixture parameter the Vinyl Chloride C2H3Cl option is selected Enter amount of substance for the gas mixture 0 to 100 0 Mol C2H4 The following condi...

Page 92: ...lected and in the Gas mixture parameter the Methane CH4 option is selected Or In the Select gas type parameter the Natural gas option is selected Enter amount of substance for the gas mixture 0 to 100...

Page 93: ...In the Select gas type parameter the Natural gas option is selected and in the Density calculation parameter the AGA Nx19 option is not selected Enter amount of substance for the gas mixture 0 to 100...

Page 94: ...the ISO 12213 2 option is selected Enter amount of substance for the gas mixture 0 to 100 0 Mol i C4H10 The following conditions are met In the Select medium parameter the Gas option is selected In t...

Page 95: ...are met In the Select medium parameter the Gas option is selected In the Select gas type parameter the Natural gas option is selected In the Density calculation parameter the ISO 12213 2 option is sel...

Page 96: ...mixture 0 to 100 0 Mol n C7H16 The following conditions are met In the Select medium parameter the Gas option is selected In the Select gas type parameter the Natural gas option is selected In the Den...

Page 97: ...medium parameter the Gas option is selected In the Select gas type parameter the Gas mixture option is selected and in the Gas mixture parameter the Oxygen O2 option is selected Or In the Select gas...

Page 98: ...10 5 2 Performing external compensation The External compensation submenu contains parameters which can be used to enter external or fixed values These values are used for internal calculations The Fi...

Page 99: ...essure Density Temperature 2nd temperature delta heat Off Atmospheric pressure In the External value parameter the Gauge pressure option is selected Enter atmospheric pressure value to be used for pre...

Page 100: ...he Steam option is selected in the Select medium parameter parameter The software options currently enabled are displayed in the Software option overview parameter Select compensation mode for steam q...

Page 101: ...essure ratings and nominal diameters DN 15 to 150 1 to 6 EN DIN ASME B16 5 Sch 40 80 Define length of the straight inlet run Dependency The unit is taken from the Length unit parameter 0 to 20 m 0 m M...

Page 102: ...vice can correct shifts in the calibration factor which are caused for example by a diameter mismatch between the device flange e g ASME B16 5 Sch 80 DN 50 2 and the mating pipe e g ASME B16 5 Sch 40...

Page 103: ...4 Configuring the pulse frequency switch output The Pulse frequency switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type...

Page 104: ...number Depends on country and nominal diameter Pulse width In the Operating mode parameter the Pulse option is selected and one of the following options is selected in the Assign pulse output paramet...

Page 105: ...ing mode parameter 103 parameter Select process variable for frequency output Off Volume flow Corrected volume flow Mass flow Flow velocity Temperature Calculated saturated steam pressure Steam qualit...

Page 106: ...f the following options is selected in the Assign frequency output parameter 105 Volume flow Corrected volume flow Mass flow Flow velocity Temperature Pressure Calculated saturated steam pressure Stea...

Page 107: ...0 Hz Failure frequency In the Operating mode parameter 103 the Frequency option is selected and one of the following options is selected in the Assign frequency output parameter 105 Volume flow Corre...

Page 108: ...vior option is selected Select diagnostic behavior for switch output Alarm Alarm or warning Warning Alarm Assign limit The Switch option is selected in the Operating mode parameter parameter The Limit...

Page 109: ...itch off point Signed floating point number Country specific 0 m h 0 ft h Switch on delay The Switch option is selected in the Operating mode parameter The Limit option is selected in the Switch outpu...

Page 110: ...cess variable parameter one of the following options is selected Volume flow Mass flow Corrected volume flow Total mass flow Condensate mass flow Energy flow Heat flow difference Control totalizer val...

Page 111: ...ation Setup menu Advanced setup Display Display Format display 112 Value 1 display 112 0 bargraph value 1 112 100 bargraph value 1 112 Decimal places 1 112 Value 2 display 112 Decimal places 2 112 Val...

Page 112: ...00 bargraph value 1 A local display is provided Enter 100 value for bar graph display Signed floating point number Depends on country and nominal diameter Decimal places 1 A measured value is specifie...

Page 113: ...ed language is preset in the device Display interval A local display is provided Set time measured values are shown on display if display alternates between values 1 to 10 s 5 s Display damping A loca...

Page 114: ...for managing the device data in the display module Cancel Execute backup Restore Duplicate Compare Clear backup data Cancel Comparison result A local display is provided Comparison between present de...

Page 115: ...vice administration purposes Navigation Setup menu Advanced setup Administration Administration Define access code Define access code 115 Confirm access code 115 Device reset 115 Parameter overview wi...

Page 116: ...tion Simulation Assign simulation process variable 117 Value process variable 117 Frequency simulation 117 Frequency value 117 Pulse simulation 117 Pulse value 117 Switch output simulation 117 Switch...

Page 117: ...at flow difference Reynolds number Enter the simulation value for the selected process variable Depends on the process variable selected 0 Frequency simulation In the Operating mode parameter the Freq...

Page 118: ...d lock 10 7 1 Write protection via access code The effects of the user specific access code are as follows Via local operation the parameters for the measuring device configuration are write protected...

Page 119: ...configuring the totalizer Language Format display Control Totalizer Contrast display Preset value Display interval 10 7 2 Write protection via write protection switch Unlike parameter write protectio...

Page 120: ...tion is displayed in the Locking status parameter In addition on the local display the symbol appears in front of the parameters in the header of the operational display and in the navigation view XXX...

Page 121: ...alculated value option For detailed information on the basic conditions for wet steam applications see the Special Documentation 10 8 2 Liquid application User specific liquid e g heat carrier oil Sel...

Page 122: ...d 2 In the Select medium parameter select the Gas option 3 In the Select gas type parameter select the Single gas option 4 In the Gas type parameter select the Methane CH4 option Configuring fluid pro...

Page 123: ...e fluid 10 In the Reference temperature parameter enter the reference temperature of the fluid Air Select medium Navigation Setup Medium selection 1 Call up the Medium selection wizard 2 In the Select...

Page 124: ...tions 9 n the Reference gross calorific value parameter enter the reference gross calorific value of the natural gas 10 In the Reference pressure parameter 86 enter the reference pressure for calculat...

Page 125: ...ic heat capacity parameter enter the heat capacity of the fluid 11 In the Z factor parameter enter the value 1 12 In the Dynamic viscosity parameter enter the viscosity of the fluid under operating co...

Page 126: ...steam water and other liquids depends on the temperature Operating density for superheated steam and all other gases depends on the temperature and process pressure Corrected volume flow calculation V...

Page 127: ...d temperature measured variables Calculation based on overheated steam until saturation point is reached Configuration of diagnostic behavior of the diagnostic message S871 Near steam saturation limit...

Page 128: ...ween warm and cold water second temperature read in via PROFIBUS PA in accordance with IAPWS IF97 ASME Vapor pressure and steam temperature The measuring device can perform the following in saturated...

Page 129: ...d value on local display PROFIBUS PA Calculation of total mass flow using steam quality and output in terms of proportions of gas and liquid Calculation of condensate mass flow using steam quality and...

Page 130: ...in the device e g data upload download reset etc Once the internal processing has been completed the parameters can be changed once again 11 2 Adjusting the operating language Detailed information To...

Page 131: ...me flow 132 Corrected volume flow 132 Mass flow 132 Flow velocity 132 Temperature 132 Calculated saturated steam pressure 132 Steam quality 132 Total mass flow 132 Condensate mass flow 132 Energy flow...

Page 132: ...eam pressure currently calculated Dependency The unit is taken from the Pressure unit parameter 75 Signed floating point number Steam quality The following conditions are met Order code for Sensor ver...

Page 133: ...ntegrated temperature measurement Displays the current value for the specific volume Dependency The unit is taken from the Specific volume unit parameter Positive floating point number Pressure One of...

Page 134: ...wing options is selected Volume flow Mass flow Corrected volume flow Total mass flow Condensate mass flow Energy flow Heat flow difference Displays the current totalizer counter value Signed floating...

Page 135: ...parameter Displays the current switch output status Open Closed 11 5 Adapting the measuring device to the process conditions The following are available for this purpose Basic settings using theSetup...

Page 136: ...tart Cancel Reset totalize Cancel Visibility depends on order options or device settings 11 7 Showing data logging The Extended HistoROM application package must be enabled in the device order option...

Page 137: ...137 Navigation Diagnostics menu Data logging Data logging Assign channel 1 138 Assign channel 2 138 Assign channel 3 138 Assign channel 4 138 Logging interval 139 Clear logging data 139 Display chann...

Page 138: ...is Gap capacity Gap capacity D Compressibility factor Electronic temperature Off Assign channel 2 The Extended HistoROM application package is available The software options currently enabled are disp...

Page 139: ...M application package is available Define the logging interval for data logging This value defines the time interval between the individual data points in the memory 1 0 to 3600 0 s 10 0 s Clear loggi...

Page 140: ...l output is within the valid range The cable of the display module is not plugged in correctly Insert the plug correctly into the main electronics module and display module Local display is dark but s...

Page 141: ...tion 119 No write access to parameters Current user role has limited access authorization 1 Check user role 55 2 Enter correct customer specific access code 55 No connection via PROFIBUS PA PROFIBUS P...

Page 142: ...stic event with the highest priority is shown Other diagnostic events that have occurred can be displayed in the Diagnostics menu Via parameter Via submenus 175 Status signals The status signals provi...

Page 143: ...generated Diagnostic information The fault can be identified using the diagnostic information The short text helps you by providing information about the fault In addition the corresponding symbol fo...

Page 144: ...n the diagnostic message Press symbol The Diagnostic list submenu opens 2 Select the desired diagnostic event with or and press The message about the remedial measures opens 3 Press simultaneously The...

Page 145: ...ubmenu 175 Status signals The status signals provide information on the state and reliability of the device by categorizing the cause of the diagnostic information diagnostic event Symbol Meaning Fail...

Page 146: ...be called up in the working area of the user interface The user is in the Diagnostics menu 1 Call up the desired parameter 2 On the right in the working area mouse over the parameter A tool tip with r...

Page 147: ...evice status is coded as per PROFIBUS PA Profile 3 02 Specification and transmitted along with the measured value to the PROFIBUS Master Class 1 via the coding byte byte 5 The coding byte is split int...

Page 148: ...Substatus Coding hex Category NE107 Alarm BAD Maintenance alarm 0x24 to 0x27 F Failure Maintenance alarm Warning GOOD Maintenance demanded 0xA8 to 0xAB M Maintenance Maintenance demanded Logbook entr...

Page 149: ...easured variables affected increase if the measuring device has one or more application packages In the case of some items of diagnostic information the diagnostic behavior can be changed Change the d...

Page 150: ...ug connections 2 Change pre amplifier 3 Change DSC sensor Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch o...

Page 151: ...Restore S Dat data 3 Change sensor Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure...

Page 152: ...lash or change main electronics module Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressu...

Page 153: ...connections 2 Change electronic modules Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pres...

Page 154: ...ice Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynolds number Specific volume...

Page 155: ...aturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynolds number Specific volume Corrected volume...

Page 156: ...ice Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynolds number Specific volume...

Page 157: ...sured variable status Quality Bad Quality substatus Maintenance alarm Coding hex 0x24 to 0x27 Status signal F Diagnostic behavior Alarm Diagnostic information Remedy instructions Influenced measured v...

Page 158: ...Change pre amplifier Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynolds numbe...

Page 159: ...tain Quality substatus Process related Coding hex 0x78 to 0x7B Status signal M Diagnostic behavior Warning 1 Diagnostic behavior can be changed This causes the overall status of the measured variable...

Page 160: ...ated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynolds number Specific volume Corrected volume flo...

Page 161: ...equency output settings Measured variable status Quality Good Quality substatus Ok Coding hex 0x80 to 0x83 Status signal S Diagnostic behavior Warning Diagnostic information Remedy instructions Influe...

Page 162: ...vior Warning Diagnostic information Remedy instructions Influenced measured variables No Short text 482 FB not Auto Cas Set Block in AUTO mode Measured variable status Quality Good Quality substatus O...

Page 163: ...iables No Short text 492 Simulation frequency output Deactivate simulation frequency output Calculated saturated steam pressure Energy flow Flow velocity Heat flow difference Low flow cut off Mass flo...

Page 164: ...ed variable status Quality Good Quality substatus Ok Coding hex 0x80 to 0x83 Status signal C Diagnostic behavior Warning Diagnostic information Remedy instructions Influenced measured variables No Sho...

Page 165: ...ic information Remedy instructions Influenced measured variables No Short text 540 Flow computer configuration incorrect Check entered reference value using the document Operating Instructions Calcula...

Page 166: ...rocess related Coding hex 0x78 to 0x7B Status signal S Diagnostic behavior Warning Diagnostic information Remedy instructions Influenced measured variables No Short text 828 Ambient temperature too lo...

Page 167: ...7B Status signal S Diagnostic behavior Warning 1 Diagnostic behavior can be changed This causes the overall status of the measured variable to change Diagnostic information Remedy instructions Influen...

Page 168: ...x7B Status signal S Diagnostic behavior Warning 1 Diagnostic behavior can be changed This causes the overall status of the measured variable to change Diagnostic information Remedy instructions Influe...

Page 169: ...B Status signal S Diagnostic behavior Warning 1 Diagnostic behavior can be changed This causes the overall status of the measured variable to change Diagnostic information Remedy instructions Influenc...

Page 170: ...y Good Quality substatus Ok Coding hex 0x80 to 0x83 Status signal S Diagnostic behavior Warning Diagnostic information Remedy instructions Influenced measured variables No Short text 844 Sensor range...

Page 171: ...saturation limit Check process conditions Calculated saturated steam pressure Density Energy flow Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynold...

Page 172: ...sure temperature 2 Check flow velocity 3 Check for flow fluctuation Calculated saturated steam pressure Density Energy flow Heat flow difference Low flow cut off Mass flow Total mass flow Switch outpu...

Page 173: ...detected Check installation Calculated saturated steam pressure Density Energy flow Flow velocity Heat flow difference Low flow cut off Mass flow Total mass flow Switch output status Pressure Reynolds...

Page 174: ...has dropped below the configured limit value for the steam quality limit value Expert System Diagnostic handling Diagnostic limits Steam quality limit Diagnostics information 873 The process temperatu...

Page 175: ...sage Previous diagnostics Two diagnostic events have already occurred Shows the diagnostic event that occurred prior to the current diagnostic event along with its diagnostic information Symbol for di...

Page 176: ...ple of the local display A maximum of 20 event messages can be displayed in chronological order If the Extended HistoROM application package order option is enabled in the device the event list can co...

Page 177: ...Write protection switch changed I1137 Electronic changed I1151 History reset I1154 Reset terminal voltage min max I1155 Reset electronic temperature I1156 Memory error trend I1157 Memory error event...

Page 178: ...ry parameter is reset to its factory setting To delivery settings Every parameter for which a customer specific default setting was ordered is reset to this customer specific value All other parameter...

Page 179: ...eplate of the sensor and transmitter in the Order code field Character string composed of letters numbers and certain punctuation marks e g Extended order code 1 Shows the 1st part of the extended ord...

Page 180: ...ackage Operation in steam simplified More robust signal processing in event of low flow rates in wet steam Operating Instructions BA01690D 06 EN 01 18 It is possible to flash the firmware to the curre...

Page 181: ...acing housing seals NOTICE When using the device in a dusty atmosphere only use the associated Endress Hauser housing seals 1 Replace defect seals only with original seals from Endress Hauser 2 The ho...

Page 182: ...e only original Endress Hauser spare parts Carry out the repair according to the Installation Instructions Observe the applicable standards federal national regulations Ex documentation XA and certifi...

Page 183: ...of services Your Endress Hauser Sales Center can provide detailed information on the services 14 4 Return The measuring device must be returned if it is need of repair or a factory calibration or if t...

Page 184: ...f the measuring device LWARNING Danger to personnel and environment from fluids that are hazardous to health Ensure that the measuring device and all cavities are free of fluid residues that are hazar...

Page 185: ...must be selected in the separate order codes Order code for measuring device feature 030 Option L or M Prepared for FHX50 display Order code for FHX50 housing feature 050 device version Option A Prepa...

Page 186: ...ce of measuring devices for industrial requirements Calculation of all the necessary data for identifying the optimum flowmeter e g nominal diameter pressure loss flow velocity and accuracy Graphic il...

Page 187: ...Instructions BA00027S and BA00059S DeviceCare Tool to connect and configure Endress Hauser field devices Innovation brochure IN01047S 15 3 System components Accessories Description Memograph M graphi...

Page 188: ...and sensor are mounted in separate locations For information on the structure of the device 12 16 3 Input Measured variable Direct measured variables Order code for Sensor version DSC sensor measurin...

Page 189: ...grated pressure temperature measurement DB Mass gas liquid 316L 316L integrated pressure temperature measurement Order code for Sensor version option mass flow integrated temperature measurement combi...

Page 190: ...tube Di is denoted in the dimensions as dimension K For detailed information see the Technical Information 217 Calculation of flow velocity v m s 4 Q m h D m i 1 3600 s h v s 4 Q min ft D ft i 1 60 s...

Page 191: ...ated without any errors Using the nominal diameter the corresponding flow can also be derived from this amplitude The minimum signal amplitude depends on the setting for the sensitivity of the DSC sen...

Page 192: ...ft i 60 s min 4 lbm 0 0624 lbm A0034304 QAmpMin Minimum measurable flow rate based on signal amplitude vAmpMin Minimum measurable flow velocity based on signal amplitude Di Internal diameter of measu...

Page 193: ...Max 1148 s D ft i 60 s min 4 lbm 0 0624 lbm A0034316 QAmpMax Maximum measurable flow rate based on signal amplitude Di Internal diameter of measuring tube corresponds to dimension K Density For gas ap...

Page 194: ...g on the operating conditions ratio between upper range value and lower range value Input signal External measured values To increase the accuracy of certain measured variables or to calculate the cor...

Page 195: ...ut Pulse width Adjustable 5 to 2000 ms Maximum pulse rate 100 Impulse s Pulse value Adjustable Assignable measured variables Mass flow Volume flow Corrected volume flow Total mass flow Energy flow Hea...

Page 196: ...50170 Volume 2 IEC 61158 2 MBP galvanically isolated Data transfer 31 25 kbit s Current consumption 10 mA Permitted supply voltage 9 to 32 V Bus connection With integrated reverse polarity protection...

Page 197: ...rotocol specific data Manufacturer ID 0x11 Ident number 0x1564 Profile version 3 02 Device description files GSD DTM DD Information and files under www endress com www profibus org Supported functions...

Page 198: ...l operation SD02 DC 1 V Option E Local operation SD03 with lighting backlighting not used DC 1 V Option E Local operation SD03 with lighting backlighting used DC 3 V Power consumption Transmitter Orde...

Page 199: ...inal discharge current 8 20 s 10 kA Temperature range 40 to 85 C 40 to 185 F 1 The voltage is reduced by the amount of the internal resistance Imin Ri Depending on the temperature class restrictions a...

Page 200: ...Min v s AmpMin D ft i 60 s min 4 lbm 0 0624 lbm A0034304 Remax Defined by internal diameter of measuring tube Mach number and maximum permitted velocity in measuring tube Re max 4 QHeigh K A0034339 Fu...

Page 201: ...ted temperature measurement external pressure compensation 2 Process pressure bar abs Flow velocity m s ft s Reynolds number Window Measured value deviation PremiumCal 3 Standard PremiumCal 3 Standard...

Page 202: ...e transmitter The overall system uncertainty which is less than 0 9 for the example above is comprised of the following measurement uncertainties uncertainty of volume flow measurement uncertainty of...

Page 203: ...ent temperature Pulse frequency output o r of reading Temperature coefficient Max 100 ppm o r 16 7 Installation Mounting requirements 20 16 8 Environment Ambient temperature range Temperature tables O...

Page 204: ...ck resistance Shock half sine according to IEC 60068 2 27 Order code for Housing option C GT20 two chamber aluminum coated compact J GT20 two chamber aluminum coated remote K GT18 two chamber 316L rem...

Page 205: ...iate standards and further information The OPL may only be applied for a limited period of time The MWP maximum working pressure for the sensors depends on the lowest rated element with regard to pres...

Page 206: ...hnical Information document Mechanical construction section Weight Compact version Weight data Including the transmitter Order code for Housing option C GT20 two chamber aluminum coated compact 1 8 kg...

Page 207: ...er code for Housing option J GT20 two chamber aluminum coated remote 2 4 kg 5 2 lb Order code for Housing option K GT18 two chamber 316L remote 6 0 kg 13 2 lb Sensor remote version Weight data Includi...

Page 208: ...Weight information in lbs DN in Weight lbs sensor connection housing Order code for Housing option J GT20 two chamber aluminum coated remote 1 sensor connection housing Order code for Housing option...

Page 209: ...0 04 25 Class 150 Class 300 0 1 40 Class 150 Class 300 0 3 50 Class 150 Class 300 0 5 80 Class 150 Class 300 1 2 1 4 100 Class 150 Class 300 2 7 150 Class 150 Class 300 6 3 7 8 200 Class 150 Class 300...

Page 210: ...lass 150 Class 300 27 0 35 0 10 Class 150 Class 300 57 0 61 0 12 Class 150 Class 300 80 0 98 0 1 ASME Materials Transmitter housing Compact version Order code for Housing option B GT18 two chamber 316...

Page 211: ...hazardous area except for XP Stainless steel 1 4404 316L Adapter for cable entry with internal thread NPT Non hazardous area and hazardous area Order code for Housing option C GT20 two chamber aluminu...

Page 212: ...ratings PN 10 16 25 40 Class 150 300 CX2MW similar to Alloy C22 2 4602 Compliant with NACE MR0175 NACE MR0103 DSC sensor Order code for Sensor version DSC sensor measuring tube option AA BA CA Pressu...

Page 213: ...high grade in the context of TA Luft clean air guidelines Housing support Stainless steel 1 4408 CF3M Screws for DSC sensor Order code for Sensor version option AA BA CA DA DB Stainless steel A2 80 ac...

Page 214: ...Two display modules are available Order code for Display Operation option C SD02 Order code for Display Operation option E SD03 A0032219 A0032221 1 Operation with pushbuttons 1 Operation with touch c...

Page 215: ...optical buttons operation possible through cover glass Display and operating elements The display and operating elements correspond to those of the display module Remote operation 56 Service interface...

Page 216: ...13359 Measurement of conductive liquid flow in closed conduits Flanged type electromagnetic flowmeters Overall length EN 61010 1 Safety requirements for electrical equipment for measurement control an...

Page 217: ...App Enter the serial number from the nameplate or scan the 2 D matrix code QR code on the nameplate Standard documentation Brief Operating Instructions Brief Operating Instructions for the sensor Mea...

Page 218: ...on the Pressure Equipment Directive SD01614D Contents Documentation code HART FOUNDATION Fieldbus PROFIBUS PA Heartbeat Technology SD02029D SD02030D SD02031D Wet steam detection SD02032D SD02033D SD02...

Page 219: ...ion preparations 33 Connection tools 29 Context menu Calling up 50 Closing 50 Explanation 50 Current consumption 198 Cyclic data transmission 62 D Declaration of Conformity 10 Define access code 118 D...

Page 220: ...protection 119 Help text Calling up 53 Closing 53 Explanation 53 HistoROM 113 I I O electronics module 12 34 Identifying the measuring device 13 Impact resistance 204 Incoming acceptance 13 Influence...

Page 221: ...4 Entering a value 54 Parameter settings Administration Submenu 115 Analog inputs Submenu 78 Communication Submenu 80 Configuration backup display Submenu 113 Data logging Submenu 136 Device informati...

Page 222: ...rt 182 Spare parts 182 Standards and guidelines 216 Status area For operational display 45 In the navigation view 47 Status signals 142 145 Storage conditions 18 Storage temperature 18 Storage tempera...

Page 223: ...ter housing Turning the transmitter housing 27 U Use of the measuring device Borderline cases 9 Incorrect use 9 see Designated use User interface Current diagnostic event 174 Previous diagnostic event...