![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634084.webp)
Operation
Settings
3-54
146iQ Instruction Manual
Thermo Scientific
●
Gas Solenoid 1
6:
Displays whether or not the Gas solenoid 1
6 is
activated. Displays electrical Current value.
●
Purge Solenoid:
Displays whether or not the purge solenoid is
activated. Displays electrical Current value.
●
Pump:
Displays whether or not the pump solenoid is activated.
Displays electrical Current value.
●
Ext/Int Solenoid:
Displays whether or not the external/internal
solenoid is activated. Displays electrical Current value.
●
Sample Reference Solenoid:
Displays whether or not the sample
reference solenoid is activated. Displays electrical Current value.
●
Permeation Solenoid 1
2:
Displays whether or not the permeation
solenoid 1
2 is activated. Displays electrical Current value.
●
Board Communication:
Displays pass/fail for communication status.
●
Power Supply:
Displays pass/fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
2.5V Diagnostic:
Displays current voltage readings. Alarm limits are
not changeable.
●
3.3V Diagnostic:
Displays current voltage readings. Alarm limits are
not changeable.
●
5V Diagnostic:
Displays current voltage readings. Alarm limits are
not changeable.
●
24V Diagnostic:
Displays current voltage readings. Alarm limits are
not changeable.
●
5V Step Board 1:
Displays current voltage readings. Alarm limits are
not changeable.
●
24V Step Board 1:
Displays current voltage readings. Alarm limits
are not changeable.
●
5V Step Board 2:
Displays current voltage readings. Alarm limits are
not changeable.
●
24V Step Board 2:
Displays current voltage readings. Alarm limits
are not changeable.
●
5V Step Board 3:
Displays current voltage readings. Alarm limits are
not changeable.
●
24V Step Board 3:
Displays current voltage readings. Alarm limits
are not changeable.
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......