![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 199](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634199.webp)
Servicing
DMC Pressure and Flow Board
Thermo Scientific
146iQ Instruction Manual
7-17
Use the following to replace the DMC pressure and flow board.
Equipment required:
Hex drive, 7/16
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug cables from the pressure and flow board.
Figure 7–20.
Flow Pressure Board, Disconnect DMC Cable
3.
Disconnect plumbing.
4.
Using 7/16 hex drive, unfasten four #6-32 socket cap head screws.
DMC Pressure
and Flow Board
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......