Operation
Calibration
Thermo Scientific
146iQ Instruction Manual
3-9
The MFC Calibration screen allows the user to calibrate the MFC’s, and
change calibration temperature and pressure.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Note
Display automatically reflects the amount and type of MFC’s
installed.
▲
Home Screen>Calibration>MFC Calibration
The MFC Calibration screen contains the following information:
●
Across:
●
Full Scale (sccm):
Shows the full-scale range of the MFC.
●
Calibration:
Allows the user to run a 2
7-point calibration for each
of the MFC’s.
●
Cal Status:
Displays Ok/Fail based on latest MFC calibration
attempt.
●
Reset Coefficients:
Resets MFC coefficients to factory default.
●
Down:
●
Zero MFC:
Calibration of zero MFC.
●
Gas 1 MFC:
Calibration of gas 1 MFC (if installed).
●
Gas 2 MFC:
Calibration of gas 2 MFC (if installed).
●
Cal Temp (C):
User sets calibration temperature.
●
Cal Pres (mmHg):
User sets calibration pressure.
MFC Calibration
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......