Calibration
Calibration of the Photometer
Thermo Scientific
146iQ Instruction Manual
4-7
The accuracy of the ozone concentrations generated in the 146iQ is
dependent on the accuracy of the internal photometer when the internal
photometer is used to calibrate the ozone generator or when it is used in
feedback control of the ozone output in normal operation. The calibration
of the photometer requires an external zero air source (ZERO) and ozone
source (SPAN), as well as an external reference photometer to validate the
zero and span concentrations. These sources should be capable of providing
the flow required by the internal photometer in addition to that of the
external reference photometer. The procedure for calibration of the internal
photometer is as follows:
1.
Connect the external ZERO/SPAN source to the bulkhead labeled
‘External Sample’ on the rear panel.
2.
Set the Internal Photometer to ‘External’ mode as shown below in
Settings>
Operation
:
3.
From the Home Screen, press Calibration>
Photometer Calibration
. It
is recommended that the averaging time for the photometer be set to a
sufficiently long value, such as 300 seconds to obtain an optimal
calibration.
4.
Ensure that the external zero air is activated, and the photometer
readings have stabilized prior to going to Calibration>Photometer
Calibration>
Calibrate Background
to perform a background
calibration of the photometer. Alternately, a manual background
calibration can also be accomplished via Calibration>Photometer
Calibration>Advanced Photometer Calibration>Manual
Calibration>
Adjust Background
(Details in the ‘
Operation
’ Chapter).
Confirm that the external reference photometer is also reading ~ 0.0
ppb.
Calibration of
the Photometer
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......