Servicing
Permeation Oven Board Replacement
Thermo Scientific
146iQ Instruction Manual
7-69
Use the following procedure to replace the permeation oven board.
Equipment required:
Phillips drive, #2
CAUTION
Allow oven to cool down prior to servicing.
▲
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug all electrical J1, J7, J8, J9 as shown.
3.
Using #2 Phillips drive, unfasten four 6-32 pan head screws.
Figure 7–77.
Replacing the Permeation Oven Board
4.
Replace permeation board and assemble in reverse order.
Permeation Oven
Board
Replacement
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......