![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634062.webp)
Operation
Data
3-32
146iQ Instruction Manual
Thermo Scientific
The Data Logging Setup screen allows the user to select data to be stored
and how it is stored.
Home Screen>Data>Advanced>Data Logging Setup
The Data Logging Setup screen contains the following information:
●
Select Data Logging Variables
: User selects instrument variables to log.
●
Period
: User selects how often data is collected by setting the duration
between logged data.
●
Data Treatment
: Toggles between Average, Current, Minimum and
Maximum. When set to average, the average value during the period
will be recorded. When set to current, the latest data will be recorded.
When set to minimum or maximum, the minimum or maximum value
during the period will be recorded.
●
Erase Data Log Records
: Allows the user to erase all values in the data log
and updates the data logging table.
Data Logging Setup
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......