Note
In the case of GPT NO
2
, the set value is the concentration of NO
calculated by the manually entered NO gas slow and zero flow.
▲
●
Gas Flow:
The user sets the MFC gas flow under the Set Value
column. Measured value displays to the right under the Measured
Value column. The lowest allowable flow setting for the MFCs is
4%.
Note
In the case of GPT NO
2
, the gas flow indicated is the flow of NO
required for GPT.
▲
●
Zero Flow:
The user sets the MFC zero flow under the Set Value
column. Measured value displays to the right under the Measured
Value column. A lower limit of 4% of the lowest MFC range is
enforced for zero flow. However, the zero MFC can be set to 0
sccm to be able to supply straight tank concentration.
●
Ozone Undiluted:
User sets the undiluted ozone concentration
under the Set Value column, if installed. Measured value displays to
the right under the Measured Value column. Undiluted ozone is
the output from the ozonator prior to dilution by the zero air.
●
Ozone Diluted:
Displays diluted ozone concentration under the Set
Value column, if installed. Measured value displays to the right
under the Measured Value column. Diluted ozone is the ozone
concentration at the output manifold.
●
Photometer:
User can choose either Off, Internal, External, or
Control, if installed.
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......