![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634135.webp)
Operation
Settings
Thermo Scientific
146iQ Instruction Manual
3-105
The Alarm Setpoints screen allows the user to view and set all settable
alarm minimum and maximum values. Alarm setpoints can also be set in
Settings>Health Check>Status and Alarms screens.
Note
User cannot set alarm limits outside of the acceptable range. The
minimum and maximum alarm limit can also be set by pressing on the
corresponding buttons located in the Settings>Health Check>Status and
Alarms screen. See “
Status and Alarms
” on page
3-39
.
▲
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>More>Instrument Settings>Alarm Setpoints
The Alarm Setpoints screen contains the following information:
●
Across:
●
Item:
Lists the items that have settable alarm limits.
●
Value:
Displays the current value for each item.
●
Low Alarm:
User sets low alarm for item.
●
High Alarm:
User sets high alarm for item.
●
Units:
Units for each item (not settable).
●
Down:
●
Photometer Bench Temperature:
Photometer bench temperature.
●
Photometer Flow:
Photometer flow alarm.
●
Photometer Pressure:
Photometer pressure alarm.
●
Photometer Cell Intensity:
Photometer cell intensity alarm.
Alarm Set Points
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......