![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 264](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634264.webp)
Optional Equipment
Connecting External Devices
9-4
146iQ Instruction Manual
Thermo Scientific
The Analog I/O Board consists of:
●
4 Isolated Analog Voltage Inputs, Input Voltage Range: 0–10 V
●
6 Isolated Analog Voltage Outputs, Three Ranges: 0–1.0 V, 0–5.0 V,
0–10 V
●
6 Isolated Analog Current Outputs, Two Ranges: 0mA–20mA, 4mA–
20mA
Table 9–3
lists the analog voltage inputs are used to monitor four external
0–10V signals.
Figure 9–4.
Analog Voltage Inputs
Table 9–3.
Analog Voltage Inputs Assignment
Terminal Number
Signal Name
1
Analog In 1
2 Analog
GND
3
Analog In 2
4 Analog
GND
5
Analog In 3
6 Analog
GND
7
Analog In 4
8 Analog
GND
Analog I/O Board
Analog Voltage Inputs
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......