Servicing
Valve Replacement
7-50
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the valves.
Equipment required:
Wrench
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug desired valve. Refer to
Figure 7–56
,
Figure 7–57
, and
Figure
7–58
.
Figure 7–56.
Valve Replacement pt 1
Valve
Replacement