![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 124](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634124.webp)
Operation
Settings
3-94
146iQ Instruction Manual
Thermo Scientific
The Manual Operation Settings screen allows the user to manually set the
gas flow, zero flow, and ozone undiluted concentration (if installed). User
will not be able to select preset span concentrations.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>More>Manual Operation>Manual Operation
Settings
The Manual Operation Settings screen contains the following information:
●
Across:
●
Manual Operation:
This column lists items associated with manual
operation.
●
Setting:
User sets up instrument operation.
●
Set Value:
Displays concentration or flow setpoints.
●
Measured Value:
Displays measured concentrations and flows. User
sets gas flow, zero flow, and undiluted ozone concentration.
●
Units:
Displays units for each item.
●
Down:
●
Gas:
Allows selection of the gas. Supports up to six external gases
(Gas 1 through 6), GPT NO
2
, O
3
(if ozonator installed),
Permeation, and Purge. The external gas names may be set
up/changed by the user via the cylinder setup screen.
●
Zero / Span:
Displays “Manual”. The set value is calculated by the
manually entered gas flow and zero flow.
Manual Operation
Settings
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......