Servicing
Sample/Ref Valve and Ext/Int Valve Replacement
7-56
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the sample/reference valve and
external/internal valve.
Equipment required:
Phillips drive, #2
Hex drive, 9/64
Adjustable wrenches, qty 2
Needle nose pliers
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug desired valve. Refer to
Figure 7–57
and
Figure 7–58
.
3.
Using two adjustable wrenches (one to hold fitting body and one to
loosen the nut), disconnect plumbing 2X.
Do not disconnect the
other 2X barb fitting with wrench
. Use needle nose pliers.
4.
Using a #2 Phillips drive, unfasten captive hardware.
Figure 7–63.
Replacing Sample/Ref Valve and Ext/Int Valve, Electrical
5.
Using a 9/64 hex drive, unfasten two 8-32 soc cap head screws.
Sample/Ref Valve
and Ext/Int Valve
Replacement
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......