Operation
Calibration
3-12
146iQ Instruction Manual
Thermo Scientific
The Manual Ozonator Calibration screen allows the user to manually
calibrate ozonator drive power to O
3
concentration.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Manual Ozonator Calibration
Home Screen>Calibration>MFC Calibration>Continue to Ozonator
Lamp Power to Concentration Screen
The Ozonator Lamp Power to Concentration Screen contains the
following information:
●
Across:
●
Ozonator Drive %:
Output power of the ozonator.
●
Select Drive:
User initiates ozonator output power setting by
pressing Start for the corresponding ozonator drive %.
●
Enter O3 Concentration (ppb):
Allows user to enter calibrated
external photometer reading for the corresponding ozonator drive
% setting.
Manual Ozonator
Calibration
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......