Servicing
Measurement Side Removal and Replacing
Thermo Scientific
146iQ Instruction Manual
7-9
Use the following procedure to remove and replace the measurement side if
necessary.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug the fan cable J18 (
Figure 7–9
).
Figure 7–9.
Unplugging the Fan Cable
3.
Unplug DMC cable (
Figure 7–10
).
Measurement
Side Removal
and Replacing