Operation
Calibration
3-10
146iQ Instruction Manual
Thermo Scientific
The Ozonator Calibration screen allows the user to run automated
calibrations for the ozonator. The automated calibration can only be
performed if the photometer option is enabled. It is recommended to
calibrate the photometer prior to proceeding to the ozonator calibration.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Ozonator Calibration
Home Screen>Calibration>MFC Calibration>Continue to Ozonator
Lamp Power to Concentration
The Continue to Ozonator Lamp Power to Concentration screen contains
the following information:
●
Across:
●
Description:
Lists items in table.
●
Lamp Power to Concentration:
User can initiate calibration by
pressing Start, and end calibration prematurely by pressing Stop.
Ozonator Calibration
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......