![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 181](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634181.webp)
Troubleshooting
Troubleshooting Guide
Thermo Scientific
146iQ Instruction Manual
6-5
Problem
Possible Cause
Action
If using a sample particulate filter,
make sure it is not blocked.
Disconnect sample particulate
filter from the sample bulkhead; if
flow increases, replace filter.
Alarm – Intensity
Defective detector
Replace detector as needed.
Defective DMC board
Replace DMC board as needed.
Alarm – Zero Check
(Option)
Instrument out of
calibration
Recalibrate instrument.
Check gas supply.
Perform a manual calibration.
Alarm – O3 Concentration
Concentration has
exceeded range limit
Check to ensure range corresponds
with expected value. If not, select
proper range.
Concentration low
Check user-defined low set point,
set to zero.
Alarm – Lamp Heater
Current (Photometer)
Defective heater
Replace heater as needed.
Alarm – Lamp Temp Sensor
Short (Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Temp Sensor
Open (Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Bench Temp
Sensor Open (Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Bench Temp
Sensor Short (Photometer)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Connection
(Photometer)
Defective lamp
Replace lamp as needed.
Alarm – Lamp Short
(Photometer)
Defective lamp
Replace lamp as needed.
Alarm – Lamp Heater
Current (Ozone)
Defective heater
Replace heater as needed.
Alarm – Bench
Temperature (Ozone)
Dirty fan filter
Clean fan filter as needed.
Alarm – Lamp Temp Sensor
Short (Ozone)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Temp Sensor
Open (Ozone)
Defective sensor
Replace sensor as needed.
Alarm – Lamp Connection
(Ozone)
Defective lamp
Replace lamp as needed.
Defective DMC board
Replace DMC board as needed.
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......