Servicing
Optional DMC Permeation Oven Solenoid Valve Replacement
7-64
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the DMC permeation oven solenoid
valve option.
Equipment required:
Phillips drive, #1
CAUTION
Allow the oven to cool down prior to servicing.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
Figure 2–1
).
2.
Unplug perm valve J5 from Step POL board 4.
Figure 7–72.
Removing and Replacing the Solenoid Valve pt 1
3.
Loosen two 4-40 screws and slide optional perm valve out.
Equipment Damage
Do not disconnect the plumbing from the valve end.
Disconnect from the attaching end only. This will prevent damaging and
leaks from the valve end.
▲
Optional DMC
Permeation Oven
Solenoid Valve
Replacement
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......