Optional Equipment
Connecting External Devices
Thermo Scientific
146iQ Instruction Manual
9-15
Table 9–7
lists the valve driver outputs.
●
Actual Output Voltage: 22
24 VDC
●
Maximum Current: 300 mA
●
Both positive and negative outputs are protected from over voltage and
over current by 500 mA fuses.
Figure 9–8.
Valve Driver Outputs
Valve Driver Outputs