Calibration
Calibration of the individual MFC outputs
4-2
146iQ Instruction Manual
Thermo Scientific
3.
From the Home Screen, press Calibration>
MFC Calibration
to see the
screen as below:
4.
Choose the calibration temperature and pressure (this should reflect
what the flow meter is using, standard or volumetric).
5.
If using measured, enter the barometric pressure and ambient
temperature as measured by the flow meter or appropriate device under
the user defined temperature and pressure buttons after disabling the
‘Standard Temp and Pressure’ option, as shown below:
6.
Choose the MFC to be calibrated under the ‘Calibration’ column, as
described in the “
Operation
” Chapter, and continue to the MFC
calibration to see the screen below:
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......