Servicing
Optional DMC Ozonator
7-46
146iQ Instruction Manual
Thermo Scientific
Use the following to replace the ozonator.
Equipment required:
Phillips drive, #2
1.
Remove the ozonator DMC from the instrument. Refer to “
DMC
Ozonator Removal
” on page
7-41
.
2.
Unplug heater and lamp cable.
Figure 7–52.
Ozonator replace pt 1
3.
Using a #2 Phillips drive, unfasten three #6-32 flat head screws.
4.
Pull out ozonator
Ozonator
Replacement