An important feature of the 146iQ is the incorporation of the
cylinder composition in the information sent down to the gas mass flow
controller(s) to better account for the variability in gas properties that may
affect their mass flow. This feature also allows a user to not have to
calibrate their gas mass flow controller(s) for a particular gas, and instead
uniformly calibrate all their flow controllers with air/N
2
and allow the gas
composition information to inform and compensate for differences in gas
properties between air/N
2
and the actual gas flowing through the flow
controller(s) at any given time. This feature will ensure accurate mass flow
outputs at all times regardless of the gas flowing through the flow
controller(s). In order for the cylinder composition to be utilized in mass
flow control and metering, it is essential that the user accurately defines
their cylinders in the Cylinder Setup screen
Additionally, in order to fully deploy the benefits of gas composition
compensation, it is important that the cylinder gas and balance gas are
chosen from the pre-populated list of gases in the Gas Name and Tank
Balance Gas columns, respectively. If the user enters a custom name, the
flow controllers will assume that the gas is Air.
▲
Home Screen>Settings>Cylinder Setup>(Gas Solenoid X)Gas
Name>Choose Solenoid X Gas Name From List
Home Screen>Settings>Cylinder Setup>(Gas Solenoid X) Tank
Balance Gas>Choose Solenoid X Balance Gas From List
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......