Servicing
Optional DMC Ozonator
7-48
146iQ Instruction Manual
Thermo Scientific
Use the following to install the Ozonator DMC.
Equipment required:
Phillips drive, #2
1.
Slide ozonator assembly downwards.
Note
Make sure ozonator plate slides into bottom pocket.
▲
Note
Make sure ozonator plate goes over the partition panel keyway
stud.
▲
Figure 7–54.
Installing the Ozonator DMC
2.
Slide ozonator assembly right.
Note
Make sure ozonator plate fits under the stud.
▲
Note
Make sure ozonator plate slides into side pocket.
▲
3.
Using a #2 Phillips drive, fasten captive hardware.
Ozonator DMC
Install
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......