The Calibrate Span Coefficient screen is used to enter the span
concentration and calibrate the O
3
coefficient. The O
3
span coefficient is
calculated, stored, and used to correct the current reading.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Photometer Calibration>Calibrate Span
Coefficient
The Calibrate Span Coefficient screen contains the following information:
●
Edit Span Concentration:
User enters the span concentration.
●
Current Concentration:
Read only. Displays current concentration.
●
Current Span Coefficient:
Read only. Displays the current span
coefficient.
●
Calculated Span Coefficient:
Read only. After the “Edit Span
Concentration” value is entered, the new calculated span coefficient is
displayed.
●
Calibrate:
When pressed, updates the coefficient and the current O
3
concentration should match the span concentration.
Calibrate Span
Coefficient
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......