Operation
Settings
Thermo Scientific
146iQ Instruction Manual
3-93
The Manual Operation screen is used to manually modify the calibrator’s
operational settings. This screen is similar to the Operation screen. When
manual operation is enabled, user can manually set gas flow, zero flow, and
ozone undiluted concentration. User will not be able to select preset span
concentrations.
Home Screen>Settings>More>Manual Operation
The Manual Operation screen contains the following information:
●
Manual Operation Settings:
When pressed, goes to the Manual
Operation screen.
●
Manual Operation:
Enables/disables manual operation. When manual
operation is enabled, the user will have access to the manual operation
screen but will not have access to the operation screen, and vice versa.
Manual Operation
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......