Servicing
Optional Photometer DMC
7-36
146iQ Instruction Manual
Thermo Scientific
Figure 7–41.
Lamp Heater Removal pt 2
3.
Slide insulation outwards.
4.
Using a 7/64 hex drive, unfasten four #6-32 cap head screws.
5.
Clean off any prior heatsink grease on the entire face.
6.
Apply a thin and even coat of heatsink grease on this entire face (in red
or shaded black).
7.
Replace lamp heater and reassemble in reverse order.