Operation
Calibration
3-8
146iQ Instruction Manual
Thermo Scientific
The Calibration screen allows the user to calibrate instrument hardware
and view calibration history. See Chapter 4 “
Calibration
” for further
instructions on how to run a calibration. If a particular option is not
installed or not being used, it is removed from the display.
Home Screen>Calibration
The Calibration screen contains the following information:
●
MFC Calibration:
The user calibrates MFCs. The calibration
temperature and calibration pressure should be chosen prior to MFC
calibration to tell the calibration algorithm whether standard or
volumetric flows are being entered by the user.
●
Ozonator Calibration:
The user calibrates the ozonator automatically.
(Visible if ozonator, photometer options are enabled.)
●
Manual Ozonator Calibration:
The user manually calibrates ozonator
with an external calibrated photometer. (Visible if ozonator option is
enabled.)
●
Photometer Calibration:
The user can calibrate the photometer. (Visible
if photometer option is enabled.)
●
Pressure Calibration:
Calibrates pressure. (Visible if ozonator and/or
photometer option(s) is/are enabled.)
●
Calibration History:
Lists all calibrations performed and calibration
checks.
Calibration
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......