Servicing
Optional Photometer DMC
7-32
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to remove the Photometer DMC and replace
the board.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug DMC cables.
3.
Disconnect 2X plumbing (
Figure 7–37
).
4.
Using a #2 Phillips drive, unfasten 4X captive hardware.
Figure 7–37.
Photometer Bench DMC Removal
Photometer DMC
Removal