![ThermoFisher Scientific 146iQ Скачать руководство пользователя страница 280](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634280.webp)
Optional Equipment
Permeation Oven
9-20
146iQ Instruction Manual
Thermo Scientific
Figure 9–12.
Installing glass tube pt 4
The computation of gas output level is shown in the following information.
Note that is assumed that all devices are properly calibrated and that all
flows are corrected to 25 °C and 1 atm.
Permeation Tube:
Q
(K)
(R)
=
(ppm)
Output
o
Where:
R = permeation rate in ng/min
Q
0
= flow rate of gas (scc/min) into the charcoal scrubber during span mode
K = constant for the specific permeant = 24.45 / MW
MW = molecular weight
Computation of
Concentrations
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......