Calibration
Calibration of the Ozone Generator
Thermo Scientific
146iQ Instruction Manual
4-5
The ozonator calibration is a 10-point calibration starting at 28% drive
power up to 90% drive power. It is recommended that all 10 points be
calibrated to achieve optimal ozone generator performance. However,
selectively desired points can be calibrated as well.
4.
Press ‘Start’ under the ‘Select Drive’ column for the desired drive % to
activate the setting (The cell turns yellow), and wait until the reference
photometer readings have stabilized before entering the concentration
in ppb under the ‘Enter O3 Concentration (ppb)’ column in the same
row (a wait time of less than 45 seconds is not accepted by the
instrument, longer wait times are recommended for a more stable
calibration). It is recommended that the external reference photometer
be set to a sufficiently long averaging time to obtain an optimal
calibration.
An alternate method to automatically calibrate the ozone generator with
the internal photometer is also available, if the internal photometer option
is available and enabled on the 146iQ. It is recommended that the internal
photometer be calibrated with a primary standard (as detailed in the
succeeding section on Photometer Calibration) prior to proceeding with
automatic ozone generator calibration. The procedure to calibrate the
ozone generator with the internal photometer is as follows:
1.
Confirm that the Photometer is set to ‘Internal’ in Settings>
Operation
as shown below:
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......