The Select Events 1
15 and 16
31 screens allow the user to select up to 31
events for each program.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Programmed Events>Select Events 1
15
The Select Events screen contains the following information:
●
Across:
●
Program Events:
This column lists event numbers.
●
Action:
User enables/disables event.
●
Duration (min):
Time duration for the event.
●
Solenoid:
User chooses gas solenoid (includes GPT NO
2
, O
3
, and
permeation “solenoids”) or off.
●
Zero/Span:
User chooses zero or span 1
6.
●
Photometer:
User chooses Internal, External, Control, or Off.
●
User D.O. 01
16:
User chooses None, Off, or On for digital
outputs.
●
Down:
●
1
15 and 16
31:
User can set up to 31 events for each program.
Select Events
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......