Troubleshooting
Troubleshooting Guide
6-4
146iQ Instruction Manual
Thermo Scientific
Problem
Possible Cause
Action
Output signal noisy
Recorder
Replace or repair recorder.
Sample is varying
Run instrument on stable ozone
source. If quite, No malfunction?
Foreign material in cell
Clean cell.
Sticky solenoid valve
Replace with known-good solenoid
valve.
Analyzer does not calibrate
properly
Leak
Check all tubing connections for
leak.
Contaminated scrubber
Perform scrubber efficiency test.
Replace if needed.
Flow/Pressure out of
calibration
Recalibrate flow/pressure. Refer to
the “
Calibration
” chapter.
Dirty system
Clean cells and flow components.
Solenoid defective
Replace with known-good solenoid
valve.
Slow response
Averaging time
Verify averaging time is set
properly.
Contaminated
photometer
Clean bench and then condition
system overnight.
Flow controller unstable
Leak
Perform a leak check.
Defective MFC
Replace mass flow controller as
needed.
No Ozone output
Defective lamp
Check for blue light when removing
ozone lamp from ozonator and
replace lamp if light is not visible.
Ozonator heater failure
Verify the ozonator is warm (>50
°
C). If not warm replace heater.
Defective DMC board
Replace DMC board as needed.
Alarm – O
3
Lamp temp
Defective lamp heater
Replace ozonator, lamp driver
board or lamp assembly.
Alarm –Lamp temp
Defective lamp heater
Replace bench lamp driver board or
lamp assembly.
Alarm – Perm Gas Temp
Incorrect Perm oven set
temperature or alarm
setting
Check that the alarm settings
match the set temperature.
Perm oven heater
thermistor or gas
thermistor error
Replace thermistor.
Perm oven failure
Replace the Perm oven.
Alarm – Flow
Flow low
Check sample capillary for
blockage. Replace as necessary.
Содержание 146iQ
Страница 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Страница 160: ......
Страница 170: ......
Страница 174: ...Maintenance Pump Rebuilding 5 4 146iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding...
Страница 176: ......
Страница 190: ...Servicing Fan Replacement 7 8 146iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan...
Страница 208: ...Servicing Power Supply Replacement 7 26 146iQ Instruction Manual Thermo Scientific Figure 7 29 Replacing Power Supply...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 146iQ Instruction Manual 7 49 Figure 7 55 Ozone DMC install pt 2...
Страница 252: ......
Страница 300: ......
Страница 301: ......